Fluke 116 CAL Product Manual - Page 5
Maintenance, Fuse Test (115
View all Fluke 116 CAL manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 5 highlights
110/113/114/115/116/117 Calibration Information Maintenance Maintenance of the Meter consists of battery and fuse replacement, as well as case cleaning. Fuse Test (115, 117) To test the fuse: 1. Set the rotary switch to Ω. 2. Plug a test lead into the v jack and touch the probe to the 10A jack, as shown in Figure 1. If the display shows a resistance value in the range of that shown in Figure 1, the fuse is good. If the display reads 0L, replace the fuse and test again. If the display shows any other value, have the Meter serviced. See How to Contact Fluke earlier in this document. Figure 1. Test the Fuse
110/113/114/115/116/117
Calibration Information
2
Maintenance
Maintenance of the Meter consists of battery and fuse replacement, as well as case cleaning.
Fuse Test (115, 117)
To test the fuse:
1.
Set the rotary switch to
Ω
.
2.
Plug a test lead into the
v
jack and touch the probe to the 10A jack, as shown in Figure 1.
If the display shows a resistance value in the range of that shown in Figure 1, the fuse is good.
If the display reads
0L
, replace the fuse and test again.
If the display shows any other value, have the Meter serviced. See
How to Contact Fluke
earlier in this
document.
Figure 1. Test the Fuse
<.5
OK
OK
X
OK
OK
Shop for Fluke products online at:
1.888.610.7664
www.
MyFlukeStore
.com