Fluke 116 CAL Product Manual - Page 13
Table 3. DMM Performance Tests: 110/114/115/116/117 cont., Display Reading
View all Fluke 116 CAL manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 13 highlights
110/113/114/115/116/117 Calibration Information Table 3. DMM Performance Tests: 110/114/115/116/117 (cont.) Display Reading[1] Step Function Range Applied 110 114 115 116 117 Set calibrator to standby, reconfigure leads, and program for amps output 31. DC Amps 10.00 A 10 A 32. AC Amps 6.000 A 5.0 A, 45 Hz 33. $ DC μAmps 600.0 μA 600 μADC 34. $ DC μAmps 600.0 μA 600 μAAC, 45 HZ 35. 36. y Temperature 37. --- Open input --- 0.0 °C --- 400 °C 38. AUTO-V 39. LoZ --- 0.5 V, 45 Hz --- 0.5 v, 0 Hz 40. 41. VoltAlert --- Hi 500 V[2], 500 Hz --- NA NA NA 9.87 to 10.13 NA 4.922 to 5.078 NA 593.8 to 606.2 9.87 to 10.13 4.922 to 5.078 NA NA 590.7 to 609.3 NA NA NA NA 0.2 to 0.8, NA AC Annunciator On 0.2 to 0.8, NA AC Annunciator On NA 489.7 to 510.3 0PEn NA -1.0 to 1.0 NA 395.0 to 405.0 NA 0.2 to 0.8, 0.2 to 0.8, NA AC AC Annunciator Annunciator On On 0.2 to 0.8, 0.2 to 0.8, NA AC AC Annunciator Annunciator On On NA 489.7 to 510.3 489.7 to 510.3 NA [3] 42. VoltAlert Lo --- NA [4] [1] If using a Fluke 9100 calibrator, the Calibrator Frequency mode must be used to obtain accurate frequency. [2] To keep from tripping the calibrator to standby, ramp up the voltage in 50 V increments with a 5 second delay between increments. [3] See steps 1-5 in Testing the VoltAlert Function (117 only). [4] See steps 6-9 in Testing the VoltAlert Function (117 only). MyFlukeStore 10 Shop for Fluke products online at: www. .com 1.888.610.7664