Fluke 116 CAL Product Manual - Page 13

Table 3. DMM Performance Tests: 110/114/115/116/117 cont., Display Reading

Page 13 highlights

110/113/114/115/116/117 Calibration Information Table 3. DMM Performance Tests: 110/114/115/116/117 (cont.) Display Reading[1] Step Function Range Applied 110 114 115 116 117 Set calibrator to standby, reconfigure leads, and program for amps output 31.  DC Amps 10.00 A 10 A 32.  AC Amps 6.000 A 5.0 A, 45 Hz 33. $ DC μAmps 600.0 μA 600 μADC 34. $ DC μAmps 600.0 μA 600 μAAC, 45 HZ 35. 36. y Temperature 37. --- Open input --- 0.0 °C --- 400 °C 38. AUTO-V 39. LoZ --- 0.5 V, 45 Hz --- 0.5 v, 0 Hz 40. 41. VoltAlert --- Hi 500 V[2], 500 Hz --- NA NA NA 9.87 to 10.13 NA 4.922 to 5.078 NA 593.8 to 606.2 9.87 to 10.13 4.922 to 5.078 NA NA 590.7 to 609.3 NA NA NA NA 0.2 to 0.8, NA AC Annunciator On 0.2 to 0.8, NA AC Annunciator On NA 489.7 to 510.3 0PEn NA -1.0 to 1.0 NA 395.0 to 405.0 NA 0.2 to 0.8, 0.2 to 0.8, NA AC AC Annunciator Annunciator On On 0.2 to 0.8, 0.2 to 0.8, NA AC AC Annunciator Annunciator On On NA 489.7 to 510.3 489.7 to 510.3 NA [3] 42. VoltAlert Lo --- NA [4] [1] If using a Fluke 9100 calibrator, the Calibrator Frequency mode must be used to obtain accurate frequency. [2] To keep from tripping the calibrator to standby, ramp up the voltage in 50 V increments with a 5 second delay between increments. [3] See steps 1-5 in Testing the VoltAlert Function (117 only). [4] See steps 6-9 in Testing the VoltAlert Function (117 only). MyFlukeStore 10 Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24

110/113/114/115/116/117
Calibration Information
10
Set calibrator to standby, reconfigure leads, and program for amps output
31.
DC Amps
10.00 A
10 A
NA
9.87 to
10.13
NA
9.87 to
10.13
32.
AC Amps
6.000 A
5.0 A, 45 Hz
NA
4.922 to
5.078
NA
4.922 to
5.078
33.
$
DC
μ
Amps
600.0
μ
A
600
μ
ADC
NA
593.8 to
606.2
NA
34.
$
DC
μ
Amps
600.0
μ
A
600
μ
AAC, 45 HZ
NA
590.7 to
609.3
NA
35.
y
Temperature
---
Open input
NA
0PEn
NA
36.
---
0.0
°
C
NA
-1.0 to 1.0
NA
37.
---
400
°
C
NA
395.0 to
405.0
NA
38.
AUTO-V
LoZ
---
0.5 V, 45 Hz
NA
0.2 to 0.8,
AC
Annunciator
On
NA
0.2 to 0.8,
AC
Annunciator
On
0.2 to 0.8,
AC
Annunciator
On
39.
---
0.5 v, 0 Hz
NA
0.2 to 0.8,
AC
Annunciator
On
NA
0.2 to 0.8,
AC
Annunciator
On
0.2 to 0.8,
AC
Annunciator
On
40.
---
500 V
[2]
, 500 Hz
NA
489.7 to
510.3
NA
489.7 to
510.3
489.7 to
510.3
41.
VoltAlert
Hi
---
NA
[3]
42.
VoltAlert
Lo
---
NA
[4]
[1]
If using a Fluke 9100 calibrator, the Calibrator Frequency mode must be used to obtain accurate frequency.
[2]
To keep from tripping the calibrator to standby, ramp up the voltage in 50 V increments with a 5 second delay between increments.
[3]
See steps 1-5 in
Testing the VoltAlert Function (117 only)
.
[4]
See steps 6-9 in
Testing the VoltAlert Function (117 only)
.
Table 3. DMM Performance Tests: 110/114/115/116/117 (cont.)
Step
Function
Range
Applied
Display Reading
[1]
110
114
115
116
117