True-rms Multimeter Performance Tests For Fluke 117 with S/N 38920290 through 42913406, use this alternate calibration procedure to verify that VoltAlert functions properly. See Figure 5. 1. Select the VoltAlert function: a. Verify that Hi is on the display. b. Verify that the beeper is silent and the red LED is off. 2. Connect wire (Pomona 4911A-12-2 or equivalent) to the High output voltage terminals of the calibrator (Fluke 5500A or equivalent). 3. Set the calibrator output to 10 V at 60 Hz. 4. Hold the Meter so that the wire contacts and runs across the top center of the unit. The wire should run perpendicular to the face of the unit (front to back of Meter, not side to side). Verify that the Meter beeper is on continuously and the red LED, at the top of the display, lights up. 5. Place the calibrator in standby mode and verify that the beeper is silent and the red LED is off. 6. Press q. Display should indicate Lo range. 7. Set the calibrator output to 30 V at 60 Hz. 8. Hold the Meter so that the wire contacts and runs across the top center of the unit. The wire should run perpendicular to the face of the unit (front to back of Meter, not side to side). Verify that the Meter beeper is on continuously and the red LED, at the top of the display, lights up. 9. Return the calibrator to standby mode and verify the Meter's beeper is silent and the red LED is off. Figure 5. Volt Alert Test (Alternate) 30.000 V OPR 60.00 Hz NORMAL V, , RTD AUX A, -SENSE, AUX V SCOPE 200V PK MAX VoltAlert HI 1000V RMS MAX LO 20V PK MAX 1V PK MAX 117 TRUE RMS MULTIMETERTC 20V RMS MAX TRIG OUT 20V PK MAX STBY OPR EARTH 7 8 9 4 5 6 1 2 3 +/ 0 • HOLD MIN MAX Lo / H RANGE OFF Hz V mV Volt Alert A A Hz A COM V 10 A FUSED MyFlukeStore Shop for Fluke products online at: www. 13 .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24

True-rms Multimeter
Performance Tests
For Fluke 117 with S/N 38920290 through 42913406, use this alternate calibration procedure to verify that
VoltAlert functions properly. See Figure 5.
1.
Select the VoltAlert function:
a.
Verify that
Hi
is on the display.
b.
Verify that the beeper is silent and the red LED is off.
2.
Connect wire (Pomona 4911A-12-2 or equivalent) to the High output voltage terminals of the calibrator
(Fluke 5500A or equivalent).
3.
Set the calibrator output to 10 V at 60 Hz.
4.
Hold the Meter so that the wire contacts and runs across the top center of the unit. The wire should run
perpendicular to the face of the unit (front to back of Meter, not side to side). Verify that the Meter beeper is
on continuously and the red LED, at the top of the display, lights up.
5.
Place the calibrator in standby mode and verify that the beeper is silent and the red LED is off.
6.
Press
q.
Display should indicate
Lo
range.
7.
Set the calibrator output to 30 V at 60 Hz.
8.
Hold the Meter so that the wire contacts and runs across the top center of the unit. The wire should run
perpendicular to the face of the unit (front to back of Meter, not side to side). Verify that the Meter beeper is
on continuously and the red LED, at the top of the display, lights up.
9.
Return the calibrator to standby mode and verify the Meter’s beeper is silent and the red LED is off.
Figure 5. Volt Alert Test (Alternate)
0
1
2
3
4
5
6
7
8
9
/
+
OPR
EARTH
20
V
PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
N
ORMAL
AUX
SCOPE
V
,
,
RTD
A,
-SE
N
SE,
AUX
V
200
V
PK
MAX
STBY
30.000
V
60.00
Hz
OPR
VoltAlert
10 A
FUSED
COM
V
A
Lo / H
OFF
Hz
A
V
Hz
A
mV
MIN MAX
RANGE
TRUE RMS MULTIMETER
117
HOLD
Volt
Alert
Shop for Fluke products online at:
1.888.610.7664
www.
MyFlukeStore
.com