Fluke 116 CAL Product Manual - Page 10

Display Test

Page 10 highlights

True-rms Multimeter Performance Tests Display Test 110/114/115/116/117 Push K and turn the rotary switch to the e position. Compare the display with the example in Figure 3. Check all segments for clarity and contrast. 113 Push K and turn the rotary switch to the k CHEK position. Compare the display with the example in Figure 3. Check all segments for clarity and contrast. Figure 3. Display Segments Backlight Test To test the backlight, press Q and verify that the backlight turns on. Keypad Test 110/114/115/116/117 To test the keypad: 1. Turn the rotary switch to . 2. Push each button separately. Each button push should cause the Meter to beep and activate a display annunciator. 3. To reset the Meter, turn the rotary switch to Off and then back to the On position. 113 To test the keypad: 1. Turn the Meter to Ω  2. Push each button separately. Each button push should cause the Meter to beep and activate a display annunciator. 3. To reset, turn off the Meter, and then back to any function position. MyFlukeStore Shop for Fluke products online at: www. 7 .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24

True-rms Multimeter
Performance Tests
Display Test
110/114/115/116/117
Push
K
and turn the rotary switch to the
e
position. Compare the display with the example in Figure 3.
Check all segments for clarity and contrast.
113
Push
K
and turn the rotary switch to the
k
CHEK
position. Compare the display with the example in
Figure 3. Check all segments for clarity and contrast.
Figure 3. Display Segments
Backlight Test
To test the backlight, press
Q
and verify that the backlight turns on.
Keypad Test
110/114/115/116/117
To test the keypad:
1.
Turn the rotary switch to
.
2.
Push each button separately.
Each button push should cause the Meter to beep and activate a display annunciator.
3.
To reset the Meter, turn the rotary switch to
Off
and then back to the
On
position.
113
To test the keypad:
1.
Turn the Meter to
Ω
2.
Push each button separately.
Each button push should cause the Meter to beep and activate a display annunciator.
3.
To reset, turn off the Meter, and then back to any function position.