Fluke 116 CAL Product Manual - Page 12
Table 3. DMM Performance Tests: 110/114/115/116/117 cont., Function, Range, Applied, Display Reading
View all Fluke 116 CAL manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 12 highlights
True-rms Multimeter Performance Tests Table 3. DMM Performance Tests: 110/114/115/116/117 (cont.) Step Function Range Applied Display Reading[1] 110 114 115 116 117 9. R 10. Continuity 600 Ω 600 Ω 20 Ω 250 Ω Beeper On Beeper Off 11. 6.000 V 5 V, 45 Hz 4.947 to 5.053 12. 6.000 V 5 V, 1 Hz 4.897 to 5.103 13. 14. AC Volts 60.00 V 60.00 V 50 V, 45 Hz 50 V, 1 Hz 49.47 to 50.53 48.97 to 51.03 15. 600.0 V 600 V, 45 Hz 593.7 to 606.3 16. 600.0 V 600 V, 1 Hz 587.7 to 612.3 17. e 6.000 V 2 V, 50 kHz[1] NA AC Volts + Hz 49.93 to 50.07 18. 6.000 V 0 V -0.002 to 0.002 19. 20. DC Volts 21. 6.000 V 60.00 V 600.0 V 5 V 50 V 600 V 4.973 to 5.027 49.73 to 50.27 596.8 to 603.2 22. 600.0 V -600 V -596.8 to -603.2 23. m 24. AC Millivolts 600.0 mV 6 mV, 45 Hz 600.0 mV 600 mV, 1 kHz 5.6 to 6.4 587.7 to 612.3 25. 26. DC Millivolts 600.0 mV 10 mV 600.0 mV 600 mV 5.6 to 6.4 587.7 to 612.3 27. Diode 2.000 V 1.9 V NA 1.881 to 1.919 28. 1000 nF Open NA 29. Capacitance 9999 μF 900 μF NA 0 to 2 881 to 919 30. LoZ Capacitance 10.00 μF 9 μF NA 7.18 to 10.82 MyFlukeStore Shop for Fluke products online at: www. 9 .com 1.888.610.7664