True-rms Multimeter Performance Tests Table 3. DMM Performance Tests: 110/114/115/116/117 (cont.) Step Function Range Applied Display Reading[1] 110 114 115 116 117 9. R 10. Continuity 600 Ω 600 Ω 20 Ω 250 Ω Beeper On Beeper Off 11. 6.000 V 5 V, 45 Hz 4.947 to 5.053 12. 6.000 V 5 V, 1 Hz 4.897 to 5.103 13.  14. AC Volts 60.00 V 60.00 V 50 V, 45 Hz 50 V, 1 Hz 49.47 to 50.53 48.97 to 51.03 15. 600.0 V 600 V, 45 Hz 593.7 to 606.3 16. 600.0 V 600 V, 1 Hz 587.7 to 612.3 17. e 6.000 V 2 V, 50 kHz[1] NA AC Volts + Hz 49.93 to 50.07 18. 6.000 V 0 V -0.002 to 0.002 19. 20.  DC Volts 21. 6.000 V 60.00 V 600.0 V 5 V 50 V 600 V 4.973 to 5.027 49.73 to 50.27 596.8 to 603.2 22. 600.0 V -600 V -596.8 to -603.2 23. m 24. AC Millivolts 600.0 mV 6 mV, 45 Hz 600.0 mV 600 mV, 1 kHz 5.6 to 6.4 587.7 to 612.3 25.  26. DC Millivolts 600.0 mV 10 mV 600.0 mV 600 mV 5.6 to 6.4 587.7 to 612.3 27.  Diode 2.000 V 1.9 V NA 1.881 to 1.919 28.  1000 nF Open NA 29. Capacitance 9999 μF 900 μF NA 0 to 2 881 to 919 30. LoZ Capacitance 10.00 μF 9 μF NA 7.18 to 10.82 MyFlukeStore Shop for Fluke products online at: www. 9 .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24

True-rms Multimeter
Performance Tests
9.
R
Continuity
600
Ω
20
Ω
Beeper On
10.
600
Ω
250
Ω
Beeper Off
11.
AC Volts
6.000 V
5 V, 45 Hz
4.947 to 5.053
12.
6.000 V
5 V, 1 Hz
4.897 to 5.103
13.
60.00 V
50 V, 45 Hz
49.47 to 50.53
14.
60.00 V
50 V, 1 Hz
48.97 to 51.03
15.
600.0 V
600 V, 45 Hz
593.7 to 606.3
16.
600.0 V
600 V, 1 Hz
587.7 to 612.3
17.
e
AC Volts + Hz
6.000 V
2 V, 50 kHz
[1]
NA
49.93 to 50.07
18.
DC Volts
6.000 V
0 V
-0.002 to 0.002
19.
6.000 V
5 V
4.973 to 5.027
20.
60.00 V
50 V
49.73 to 50.27
21.
600.0 V
600 V
596.8 to 603.2
22.
600.0 V
-600 V
-596.8 to -603.2
23.
m
AC Millivolts
600.0 mV
6 mV, 45 Hz
5.6 to 6.4
24.
600.0 mV
600 mV, 1 kHz
587.7 to 612.3
25.
DC Millivolts
600.0 mV
10 mV
5.6 to 6.4
26.
600.0 mV
600 mV
587.7 to 612.3
27.
Diode
2.000 V
1.9 V
NA
1.881 to 1.919
28.
Capacitance
1000 nF
Open
NA
0 to 2
29.
9999
μ
F
900
μ
F
NA
881 to 919
30.
LoZ
Capacitance
10.00
μ
F
9
μ
F
NA
7.18 to 10.82
Table 3. DMM Performance Tests: 110/114/115/116/117 (cont.)