Fluke 116 CAL Product Manual - Page 9
Performance Tests, Required Equipment
View all Fluke 116 CAL manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 9 highlights
110/113/114/115/116/117 Calibration Information Performance Tests XW Warning To prevent electric shock, do not perform the performance test procedures unless the Meter is fully assembled. The following performance tests verify the complete operation of the Meter and check the accuracy of each Meter function against its specifications. The recommended calibration interval is 12 months. If the Meter fails any part of the test, calibration adjustment and/or repair is indicated. In the performance tests, the Meter is referred to as the device under test (DUT). See Table 1 for a list of the features for each Meter. Required Equipment Table 2 is a list of the equipment required to conduct a performance test on the Meter. Table 2. Equipment Requirements Equipment 5522 Multi-product Calibrator (or equivalent) Fluke 80 AK K-type Thermocouple Adapter Accessory K-type Thermocouple, mini-plug on both ends Double Banana plug Measurement Function DC Volts DC Current AC Volts AC Current Resistance Capacitance Temperature Frequency Temperature Temperature VoltAlert [1] For accuracy specifications, see the Users Manual for the product. Range[1] 10 mV to 600 V 600 μA to 10 A 6 mV to 600 V 600 μA to 10 A 0 to 30 MΩ 9 to 900 μF 0 °C to 400 °C 2 V, 50 kHz NA NA NA MyFlukeStore 6 Shop for Fluke products online at: www. .com 1.888.610.7664