Fluke 116 CAL Product Manual - Page 18
Restoring the Default Password, Warning, To prevent electric shock or personal injury
View all Fluke 116 CAL manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 18 highlights
True-rms Multimeter Calibration Adjustment Restoring the Default Password If the calibration password is forgotten, the default password (1234) can be manually restored: XW Warning To prevent electric shock or personal injury, remove the test leads and any input signal before removing the Meter's back case. 1. Remove the Meter back case. Leave the PCA in the top case. 2. Apply 9.0 V across the battery contacts (XBT1) + and (XBT2) - on the back of the PCA. See Figure 6. 3. Turn the rotary switch from OFF to any ON position. 4. Short across the S7 CAL keypad on the back of the PCA. See Figure 6. The Meter should beep. The default password is now restored. 5. Remove the 9.0 V supply and replace the Meter back case. Figure 6. Calibration Password Reset Short S7 to reset to Default password XBT2 XBT1 MyFlukeStore Shop for Fluke products online at: www. 15 .com 1.888.610.7664