Invacare TDXSR-CG-HD Owners Manual - Page 102

DO NOT remove fuse or mounting hardware from POSITIVE + RED battery cable mounting screw.

Page 102 highlights

7 SETUP/MAINTENANCE 7.19 Connecting/Disconnecting the Battery Wiring Harness - TDX SP and TDX SR with Elevating Seat Only ƽ WARNING The use of rubber gloves is recommended when working with batteries. Invacare strongly recommends that battery installation and battery replacement ALWAYS be done by a qualified technician. After ANY adjustments, repair or service and before use, make sure all attaching hardware is tightened securely otherwise injury or damage may result. 22NF batteries weigh 37 pounds each. Use proper lifting techniques (lift with your legs) to avoid injury. ALWAYS use a battery lifting strap when lifting a battery. It is the most convenient method and assures that the battery acid will not spill. It also helps to prolong the life of the battery. DO NOT tip the batteries. Keep the batteries in an upright position. NEVER allow any of your tools and/or battery cable(s) to contact BOTH battery post(s) at the same time. An electrical short may occur and serious personal injury or damage may occur. When tightening the clamps, ALWAYS use a box wrench. Pliers will "round off" the nuts. NEVER wiggle the battery terminal(s)/post(s) when tightening. The battery may become damaged. The POSITIVE (+) RED battery cable MUST connect to the POSITIVE (+) battery terminal(s)/post(s), otherwise serious damage will occur to the electrical system. Install protective caps on POSITIVE (+) and NEGATIVE (-) battery terminals. DO NOT remove fuse or mounting hardware from POSITIVE (+) RED battery cable mounting screw. Unless otherwise indicated, make sure power to the wheelchair is OFF before performing these procedures. After ANY adjustments, repair or service and before use, make sure all attaching hardware is tightened securely otherwise injury or damage may occur.  If there is battery acid in the bottom of the battery tray or on the sides of the battery(ies), apply baking soda to these areas to neutralize the battery acid. Before reinstalling the existing or new battery(ies), clean the baking soda from the battery tray or battery(ies) being sure to avoid contact with skin and eyes. Determine source of contamination. NEVER install/ reinstall a battery with a cracked or otherwise damaged case. Invacare® Formula™ CG Seating 102 Part No 1143155

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124

7
SETUP/MAINTENANCE
Invacare® Formula™ CG Seating
102
Part No 1143155
7.19
Connecting/Disconnecting the Battery Wiring Harness - TDX SP and TDX SR with Elevating Seat Only
±
WARNING
The use of rubber gloves is recommended when working with batteries.
Invacare strongly recommends that battery installation and battery replacement ALWAYS be done by a qualified technician.
After ANY adjustments, repair or service and before use, make sure all attaching hardware is tightened securely -
otherwise injury or damage may result.
22NF batteries weigh 37 pounds each. Use proper lifting techniques (lift with your legs) to avoid injury.
ALWAYS use a battery lifting strap when lifting a battery. It is the most convenient method and assures that the battery
acid will not spill. It also helps to prolong the life of the battery.
DO NOT tip the batteries. Keep the batteries in an upright position.
NEVER allow any of your tools and/or battery cable(s) to contact BOTH battery post(s) at the same time. An electrical
short may occur and serious personal injury or damage may occur.
When tightening the clamps, ALWAYS use a box wrench. Pliers will “round off” the nuts. NEVER wiggle the battery
terminal(s)/post(s) when tightening. The battery may become damaged.
The POSITIVE (+) RED battery cable MUST connect to the POSITIVE (+) battery terminal(s)/post(s), otherwise serious
damage will occur to the electrical system.
Install protective caps on POSITIVE (+) and NEGATIVE (-) battery terminals.
DO NOT remove fuse or mounting hardware from POSITIVE (+) RED battery cable mounting screw.
Unless otherwise indicated, make sure power to the wheelchair is OFF before performing these procedures.
After ANY adjustments, repair or service and before use, make sure all attaching hardware is tightened securely -
otherwise injury or damage may occur.
If there is battery acid in the bottom of the battery tray or on the sides of the battery(ies), apply baking soda to these areas
to neutralize the battery acid. Before reinstalling the existing or new battery(ies), clean the baking soda from the battery
tray or battery(ies) being sure to avoid contact with skin and eyes. Determine source of contamination. NEVER install/
reinstall a battery with a cracked or otherwise damaged case.