Invacare TDXSR-CG-HD Owners Manual - Page 117

Service Indicator Light

Page 117 highlights

Service Indicator Light Diagnostics  For this procedure, refer to FIGURE 1. 8 TROUBLESHOOTING Service Indicator Light FIGURE 1 Service Indicator Light Diagnostics NUMBER OF FLASHES 1 2 3 4 5 6 ERROR CODE DESCRIPTION User Fault Battery Fault Left Motor Fault Right Motor Fault Left Park Brake Fault Right Park Brake Fault POSSIBLE SOLUTION Release joystick to neutral and try again. Charge the batteries. Refer to the wheelchair base owner's manual. Check that battery cables are connected properly. If necessary, replace batteries. Refer to the wheelchair base owner's manual.. Contact Invacare/Dealer for service. Contact Invacare/Dealer for service. Ensure brake lever is in the drive position before turning on the wheelchair. Ensure motor cable is plugged into the controller. Contact Invacare/Dealer for service. Ensure brake lever is in the drive position before turning on the wheelchair. Ensure motor cable is plugged into the controller. Contact Invacare/Dealer for service. Part No 1143155 117 Invacare® Formula™ CG Seating

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124

8
TROUBLESHOOTING
Part No 1143155
117
Invacare® Formula™ CG Seating
Service Indicator Light Diagnostics
FIGURE 1
Service Indicator Light Diagnostics
For this procedure, refer to FIGURE 1.
NUMBER OF
FLASHES
ERROR CODE
DESCRIPTION
POSSIBLE SOLUTION
1
User Fault
Release joystick to neutral and try again.
2
Battery Fault
Charge the batteries. Refer to the wheelchair base owner’s manual.
Check that battery cables are connected properly. If necessary, replace batteries. Refer to the
wheelchair base owner’s manual..
3
Left Motor Fault
Contact Invacare/Dealer for service.
4
Right Motor Fault
Contact Invacare/Dealer for service.
5
Left Park Brake Fault
Ensure brake lever is in the drive position before turning on the wheelchair. Ensure motor cable is
plugged into the controller.
Contact Invacare/Dealer for service.
6
Right Park Brake Fault
Ensure brake lever is in the drive position before turning on the wheelchair. Ensure motor cable is
plugged into the controller.
Contact Invacare/Dealer for service.
Service Indicator Light