Invacare TDXSR-CG-HD Owners Manual - Page 118

Probable Cause

Page 118 highlights

8 TROUBLESHOOTING NUMBER OF FLASHES 7 8 9 10 11 ERROR CODE DESCRIPTION Remote Fault Controller Fault Communications Fault General Fault Incompatible or incorrect Remote POSSIBLE SOLUTION Check to make sure joystick is connected properly. Turn Joystick off then on. Contact Invacare/Dealer for service. Contact Invacare/Dealer for service. Check joystick cable connections. Check joystick cable and connectors for damage. Contact Invacare/Dealer for service. Contact Invacare/Dealer for service. Wrong type of remote connected. Contact Invacare/Dealer for service. CMPJ™+, PSR+, PSF+ Joysticks or Displays SYMPTOM PROBABLE CAUSE SOLUTIONS LEFT BRAKE FAULT or RIGHT BRAKE FAULT displays and wheelchair does not drive. Motor lock levers disengaged (Error Engage motor lock levers. Refer to owner's manual shiped code E9 or E10). with wheelchair. CHARGER PLUGGED IN displays. Battery charger connected (Error Unplug battery charger from the wheelchair. Refer to code E28). owner's manual shiped with wheelchair. BATTERY FAULT displays and the wheelchair does not drive. Batteries need to be charged (Error code E14). Charge batteries. Refer to owner's manual shiped with wheelchair. If batteries fail to charge properly, check battery charger or replace batteries. Refer to About Batteries on page 92. JOYSTICK TIMEOUT displays and the wheelchair does not drive. Joystick or input device is disconnected (Error code 32). Turn off power, reconnect the joystick of input device and turn power on. Invacare® Formula™ CG Seating 118 Part No 1143155

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124

8
TROUBLESHOOTING
Invacare® Formula™ CG Seating
118
Part No 1143155
CMPJ™+, PSR+, PSF+ Joysticks or Displays
7
Remote Fault
Check to make sure joystick is connected properly. Turn Joystick off then on.
Contact Invacare/Dealer for service.
8
Controller Fault
Contact Invacare/Dealer for service.
9
Communications Fault
Check joystick cable connections. Check joystick cable and connectors for damage.
Contact Invacare/Dealer for service.
10
General Fault
Contact Invacare/Dealer for service.
11
Incompatible or
incorrect Remote
Wrong type of remote connected. Contact Invacare/Dealer for service.
SYMPTOM
PROBABLE CAUSE
SOLUTIONS
LEFT BRAKE FAULT or RIGHT BRAKE
FAULT displays and wheelchair does not
drive.
Motor lock levers disengaged (Error
code E9 or E10).
Engage motor lock levers. Refer to owner’s manual shiped
with wheelchair.
CHARGER PLUGGED IN displays.
Battery charger connected (Error
code E28).
Unplug battery charger from the wheelchair. Refer to
owner’s manual shiped with wheelchair.
BATTERY FAULT displays and the wheelchair
does not drive.
Batteries need to be charged (Error
code E14).
Charge batteries. Refer to owner’s manual shiped with
wheelchair. If batteries fail to charge properly, check battery
charger or replace batteries. Refer to About Batteries
on
page 92.
JOYSTICK TIMEOUT displays and the
wheelchair does not drive.
Joystick or input device is
disconnected (Error code 32).
Turn off power, reconnect the joystick of input device and
turn power on.
NUMBER OF
FLASHES
ERROR CODE
DESCRIPTION
POSSIBLE SOLUTION