DOCUSP 3.74 5. On the DocuSP interface, select Setup and System Preferences. Select the Job Policy tab. 6. Ensure that the setting for Retain PDL files of last 50 jobs is set as enabled. 7. Select [OK]. Do not restart the system if you are prompted to do so. 8. In the background, right-click to open the menu. Open a Terminal window. 9. Login as root user. 10. Type: /opt/XRXnps/bin/setretainPDL # and press , where # is a number between 1 and 100. A value of 0 maintains the current setting. 11. Close the Terminal window. Restart the DocuSP software. Reset the Xrite Spectrophotometer Perform the following to reset the spectrophotometer: 1. Ensure that the calibration strip that matches the serial number on the spectrophotometer is available. If not, order a new one before continuing. 2. On the DocuSP controller background menu, open a Terminal window. 3. At the % prompt, type: cd /opt/XRXnps/XRXcalLib/bin 4. From the % prompt, type: ./resetDTP This will restore the spectrophotometer to the factory default settings. 5. After the reset is complete, follow the standard procedure to calibrate the spectrophotometer using the Xrite calibration strip. 5-2 XEROX DOCUMENT SERVICES PLATFORM SERIES GETTING STARTED

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54

DOCUSP 3.74
5-2
XEROX DOCUMENT SERVICES PLATFORM SERIES
GETTING STARTED
5.
On the DocuSP interface, select Setup and System
Preferences.
Select the Job Policy tab.
6.
Ensure that the setting for Retain PDL files of last 50 jobs is
set as enabled.
7. Select [OK].
Do not restart the system if you are prompted
to do so.
8.
In the background, right-click to open the menu.
Open a
Terminal window.
9.
Login as root user.
10. Type: /opt/XRXnps/bin/setretainPDL # and press <Enter>,
where # is a number between 1 and 100.
A value of 0
maintains the current setting.
11. Close the Terminal window.
Restart the DocuSP software.
Reset the Xrite Spectrophotometer
Perform the following to reset the spectrophotometer:
1.
Ensure that the calibration strip that matches the serial
number on the spectrophotometer is available.
If not,
order a new one before continuing.
2.
On the DocuSP controller background menu, open a
Terminal window.
3.
At the % prompt, type:
cd /opt/XRXnps/XRXcalLib/bin
4.
From the % prompt, type:
./resetDTP
This will restore the spectrophotometer to the factory
default settings.
5.
After the reset is complete, follow the standard
procedure to calibrate the spectrophotometer using the
Xrite calibration strip.