Fluke CNX v3000 Manual - Page 14
MIN MAX AVG Record Mode, To pause MIN MAX AVG record, push
View all Fluke CNX v3000 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 14 highlights
CNX 3000 Users Manual MIN MAX AVG Record Mode The MIN MAX AVG record mode records the minimum and maximum input values, and calculates a running average of all measurements. The Product beeps when a new high or low is sensed. Note For dc functions, accuracy is the specified accuracy of the measurement function, ±12 counts for changes longer than 250 ms in duration. For ac functions, accuracy is the specified accuracy of the measurement function ±40 counts for changes longer than 900 ms in duration. To start a MIN MAX AVG record session: 1. Make sure the Product is set to the correct measurement function and on the correct range. Autorange is disabled while in a MIN MAX AVG record session. 2. Push . and Max show at the top of the display. The measurement in the display is the maximum value measured. It will change only when a new maximum value is sensed. 3. To pause MIN MAX AVG record, push . shows in the display while record is paused. Recorded values are not deleted. To continue record session, push . 4. To exit and erase the MIN, MAX, and AVG values, push for 1 second or turn the rotary switch. 5. To see the other recorded values (minimum and average), push . Each push of the button shows a different recorded value. The value shown in the display is identified with Max, Min, or Avg to the right of the MIN MAX icon. Note Battery save or sleep mode is turned off in MIN MAX AVG record mode. 6 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412