Fluke CNX v3000 Manual - Page 50

Input Characteristics, Typical Short Circuit Current

Page 50 highlights

CNX 3000 Users Manual Input Characteristics Function Overload Protection Input Impedance (nominal) Common Mode Rejection Ratio (1 kΩ unbalance) Normal Mode Rejection  1100 V rms >10 MΩ 120 dB at dc, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz  1100 V rms >10 MΩ < 100 pF > 60 dB, dc to 60 Hz  1100 V rms >10 MΩ 120 dB at dc, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz 1100 V rms 1100 V rms Open Circuit Test Voltage

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50

CNX 3000
Users Manual
42
Input Characteristics
Function
Overload
Protection
Input Impedance
(nominal)
Common Mode
Rejection Ratio
(1 k
Ω
unbalance)
Normal Mode Rejection
1100 V rms
>10 M
Ω
<100 pF
> 120 dB at dc, 50 Hz or 60 Hz
> 60 dB at 50 Hz or 60 Hz
1100 V rms
>10 M
Ω
< 100 pF
> 60 dB, dc to 60 Hz
1100 V rms
>10 M
Ω
<100 pF
> 120 dB at dc, 50 Hz or 60 Hz
> 60 dB at 50 Hz or 60 Hz
Open Circuit Test
Voltage
Full Scale Voltage
Typical Short Circuit Current
To 6 M
Ω
50 M
Ω
Ω
/
1100 V rms
<2.7 V dc
<0.7 V dc
<0.9 V dc
<350
μ
A
/
1100 V rms
<2.7 V dc
2.000 V dc
<1.1 mA
Function
Overload Protection
Overload
mA
Fused, 44/100 A, 1000 V FAST Fuse
600 mA overload for 2 minutes maximum, 10 minutes rest minimum
MIN MAX Recording
Function
Accuracy
DC Functions
The specified accuracy of the measurement function
±
12 counts for changes >350 ms in duration.
AC Functions
The specified accuracy of the measurement function
±
40 counts for changes >900 ms in duration.
Shop for Fluke products online at: