Invacare CLTD Owners Manual - Page 34
Adjusting Adjustable Angle Flip-up Footplates, Depth Adjustment, Angle Adjustment
View all Invacare CLTD manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 34 highlights
SECTION 4-FRONT RIGGINGS Adjusting Adjustable Angle Flip-up Footplates ƽ WARNING When determining the angle of the footplates, make sure the rear of the footplates do not interfere with the movement of the front casters. Otherwise, injury or damage may occur. NOTE: For this procedure, refer to FIGURE 4.7 Depth Adjustment 1. Remove the two flat screws and locknuts that secure footplate to the half clamp. NOTE: Observe the angle of the footplate for reinstallation. 2. Move the footplate to one of four mounting positions. NOTE: If desired depth is still not obtained, rotate the half clamp on the footplate hinge 180°. 3. Retighten the two flat screws and locknuts. NOTE: The setting for positioning the footplate on the half clamp may vary for each footplate. 4. Repeat STEPS 1-3 for opposite footplate, if desired. Angle Adjustment 1. Loosen, but DO NOT remove the adjustment screw in the half clamp. 2. Position the footplate to the necessary angle to accommodate the user (Detail "A" of FIGURE 8). 3. Retighten the adjustment screw. Torque to over 90 in.-lbs., but no more than 300 in.-lbs. 4. Repeat STEPS 1-3 for opposite footplate, if desired. Flat Screws Footplate Half Clamp Adjustment Screw 90° Footrest Support DETAIL "A" - SIDE VIEW OF FOOTPLATE AND FOOTREST SUPPORT Footrest Support Footplate Hinge Footplate Locknuts FIGURE 4.7 Adjusting Adjustable Angle Flip-up Footplates Compass™SPT™ 34 Part No 1122134