Invacare SPT Owners Manual - Page 44
Adjusting T-Arm Width,
View all Invacare SPT manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 44 highlights
SECTION 5-ARMS Adjusting T-Arm Width NOTE: For this procedure, refer to FIGURE 5.5. 1. Remove the two mounting screws that secure the arm pad to the arm tube. 2. Turn the arm pad around and reposition the arm pad on the arm tube. 3. Re-secure the arm pad to the arm tube with the two mounting screws. Tighten securely. 4. Repeat for the opposite side, if necessary. Arm Pad Mounting Screw Arm Tube Mounting Screw FIGURE 5.5 Adjusting T-Arm Width Adjusting T-Arm Depth NOTE: For this procedure, refer to FIGURE 5.6. 1. Remove the two mounting screws that secure the arm pad to the arm tube. 2. Remove the two socket screws that secure the arm tube to the T-Arm post. 3. Reposition the arm tube on the T-Arm post: A. Desk Length Arms - to one of three positions depending on the desired arm pad depth. B. Full Length Arms - to one of five positions depending on the desired arm pad depth. NOTE: Two additional positions are obtainable by turning the arm tube 180°. 4. Re-secure the arm tube to the T-Arm post with the two socket screws. Torque to 60-70 in.-lbs. 5. Reattach the arm pad to the arm tube with the two mounting screws. Tighten securely. Arm Pad Socket Screws Arm Tube Mounting Screw T-Arm Post Mounting Screw FIGURE 5.6 Adjusting T-Arm Depth 6. Repeat for the opposite side, if necessary. Compass™SPT™ 44 Part No 1122134