Fluke 123B/S Service Manual - Page 69
Input B Trigger Sensitivity Test, to open the SCOPE INPUTS menu.
View all Fluke 123B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 69 highlights
4 Performance Verification 4.5 Input A and Input B Tests INPUT: I B | SCREEN UPDATE: I FREE RUN | AUTO RANGE: I >15HZ 3. Set the 5500A to source a leveled sine wave of 1.2V peak-to-peak, 50 kHz (SCOPE output, MODE levsin). 4. Adjust the amplitude of the sine wave to 6 divisions on the test tool display. 5. Set the 5500A to 20 MHz (Fluke 123) or 40 MHz (Fluke 124), without changing the amplitude. 6. Observe the Input B trace check to see if it is ≥ 4.2 divisions. 7. When you are finished, set the 5500A to Standby. Note The lower transition point is tested in Section 4.5.11. 4.5.7 Input B Trigger Sensitivity Test Proceed as follows to test the Input B trigger sensitivity: 1. Connect the test tool to the 5500A as for the previous test (see Figure 4-4). 2. Select the following test tool setup: • Turn Input B on (if not already on). • Press to select auto ranging (AUTO in top of display). Do not press anymore! • Using change the sensitivity to select manual sensitivity ranging, and lock the Input B sensitivity on 200 mV/div. • Press • Press to open the SCOPE INPUTS menu. to open the TRIGGER menu, and choose: INPUT: I B | SCREEN UPDATE: I FREE RUN | AUTO RANGE: I >15HZ 3. Set the 5500A to source a 5 MHz leveled sine wave of 100 mV peak-to-peak (SCOPE output, MODE levsin). 4. Adjust the amplitude of the sine wave to 0.5 division on the display. 5. Verify that the signal is well triggered. If it is not, press to enable the up/down arrow keys for Trigger Level adjustment; adjust the trigger level and verify that the signal will be triggered now. The trigger level is indicated by the trigger icon ( ). 6. Set the 5500A to source a 25 MHz (Fluke 123) or 40 MHz (Fluke 124) leveled sine wave of 400 mV peak-to-peak. 7. Adjust the amplitude of the sine wave 1.5 divisions on the test tool display. 8. Verify that the signal is well triggered. If it is not, press to enable the up/down arrow keys for Trigger Level adjustment; adjust the trigger level and verify that the signal will be triggered now. 9. Set the 5500A to source a 40 MHz (Fluke 123) or 60 MHz (Fluke 124) leveled sine wave of 1.8V peak-to-peak. 10. Adjust the amplitude of the sine wave to exactly 4 divisions on the test tool display. 4-11