Fluke 123B/S Service Manual - Page 78
Input A and B High Voltage AC/DC Accuracy Test
View all Fluke 123B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 78 highlights
123/124 Service Manual 4.5.14 Input A and B High Voltage AC/DC Accuracy Test Warning Dangerous voltages will be present on the calibration source and connecting cables during the following steps. Ensure that the calibrator is in standby mode before making any connection between the calibrator and the test tool. Proceed as follows to test the Input A&B High Voltage AC and DC Accuracy: 1. Connect the test tool to the 5500A as shown in Figure 4-6. Figure 4-6. Test Tool Input A-B to 5500A Normal Output for >300V ST8129.WMF 2. Select the following test tool setup: • Press to select auto ranging (AUTO in top of display). Do not press anymore! • Press to open the INPUT A MEASUREMENTS menu, and choose: MEASURE on A: I VAC • Press to open the INPUT A MEASUREMENTS menu, and choose: MEASURE on A: I VDC (VDC becomes main reading, VAC secondary reading) • Press to open the INPUT B MEASUREMENTS menu, and choose: INPUT B: I ON | MEASURE on B: I VAC • Press to open the INPUT B MEASUREMENTS menu, and choose: INPUT B: I ON | MEASURE on B: I VDC • Move the Input A and Input B ground level (indicated by zero icon ) to the center grid line. Proceed as follows: Press to enable the arrow keys for moving the Input A ground level. Press to enable the arrow keys for moving the Input B ground level. Using the keys move the ground level. 4-20