Fluke 123B/S Service Manual - Page 94
Continue at READY, Table 5-1. HF Gain Calibration Points Fast, Cal step, 5500A Setting
View all Fluke 123B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 94 highlights
123/124 Service Manual 3. Set the 5500A to source a 1 kHz fast rising edge square wave (Output SCOPE, MODE edge) to the first calibration point in Table 5-1. 4. Set the 5500A in operate (OPR). 5. Press to start the calibration. 6. Wait until the display shows calibration status READY . 7. Press to select the next calibration step, set the 5500A to the next calibration point, and start the calibration. Continue through all calibration points in Table 5-1. 8. Set the 5500A to source a 1 kHz square wave (Output SCOPE, MODE wavegen, WAVE square), to the first calibration point in Table 5-2. 9. Press to select the first step in Table 5-2. 10. Press to start the calibration. 11. Wait until the display shows calibration status READY. 12. Press to select the next calibration step, set the 5500A to the next calibration point, and start the calibration. Continue through all calibration points Table 5-2. 13. When you are finished, set the 5500A to Standby. 14. Continue at Section 5.6.2. Table 5-1. HF Gain Calibration Points Fast Cal step 5500A Setting 1) (1 kHz, no 50Ω!) Test Tool Input Signal Requirements 1) (1 kHz, trise