Fluke 124 FE 123 & 124 Users Manual - Page 71
Horizontal, Trigger, Sensitivity A and B Fluke 123
View all Fluke 124 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 71 highlights
Horizontal Scope Modes Normal, Single, Roll Ranges Normal: equivalent sampling (Fluke 123) ..... 20 ns to 500 ns/div equivalent sampling (Fluke 124) ..... 10 ns to 500 ns/div real time sampling 1 µs to 5 s/div Single (real time 1 µs to 5 s/div Roll (real time 1s to 60 s/div Sampling Rate (for both channels simultaneously) Equivalent sampling (repetitive signals up to 1.25 GS/s Real time sampling: 1 µs to 5 ms/div 25 MS/s 10 ms to 60 s/div 5 MS/s Time Base Accuracy Equivalent sampling 0.4% +0.04 time/div) Real time sampling 0.1% +0.04 time/div) Glitch Detection 40 ns @ 20 ns to 5 ms/div ≥200 ns @ 10 ms to 60 s/div Glitch detection is always active. Horizontal Move 10 divisions Trigger point can be positioned anywhere across the screen. 4 Specifications Dual Input Oscilloscope Trigger Screen Update Free Run, On Trigger Source A, B, EXT EXTernal via optically isolated trigger probe ITP120 (optional accessory) Sensitivity A and B (Fluke 123) @ DC to 5 MHz 0.5 divisions or 5 mV @ 25 MHz 1.5 divisions @ 40 MHz 4 divisions Sensitivity A and B (Fluke 124) @ DC to 5 MHz 0.5 divisions or 5 mV @ 40 MHz 1.5 divisions @ 60 MHz 4 divisions Slope Positive, Negative Video on A interlaced video signals only Modes Lines, Line Select Standards PAL, NTSC, PAL+, SECAM Polarity Positive, Negative Sensitivity 0.6 divisions sync. 63