Fluke 124 FE 123 & 124 Users Manual - Page 71

Horizontal, Trigger, Sensitivity A and B Fluke 123

Page 71 highlights

Horizontal Scope Modes Normal, Single, Roll Ranges Normal: equivalent sampling (Fluke 123) ..... 20 ns to 500 ns/div equivalent sampling (Fluke 124) ..... 10 ns to 500 ns/div real time sampling 1 µs to 5 s/div Single (real time 1 µs to 5 s/div Roll (real time 1s to 60 s/div Sampling Rate (for both channels simultaneously) Equivalent sampling (repetitive signals up to 1.25 GS/s Real time sampling: 1 µs to 5 ms/div 25 MS/s 10 ms to 60 s/div 5 MS/s Time Base Accuracy Equivalent sampling 0.4% +0.04 time/div) Real time sampling 0.1% +0.04 time/div) Glitch Detection 40 ns @ 20 ns to 5 ms/div ≥200 ns @ 10 ms to 60 s/div Glitch detection is always active. Horizontal Move 10 divisions Trigger point can be positioned anywhere across the screen. 4 Specifications Dual Input Oscilloscope Trigger Screen Update Free Run, On Trigger Source A, B, EXT EXTernal via optically isolated trigger probe ITP120 (optional accessory) Sensitivity A and B (Fluke 123) @ DC to 5 MHz 0.5 divisions or 5 mV @ 25 MHz 1.5 divisions @ 40 MHz 4 divisions Sensitivity A and B (Fluke 124) @ DC to 5 MHz 0.5 divisions or 5 mV @ 40 MHz 1.5 divisions @ 60 MHz 4 divisions Slope Positive, Negative Video on A interlaced video signals only Modes Lines, Line Select Standards PAL, NTSC, PAL+, SECAM Polarity Positive, Negative Sensitivity 0.6 divisions sync. 63

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86

Specifications
Dual Input Oscilloscope
4
63
Horizontal
Scope Modes
...................................
Normal, Single, Roll
Ranges
Normal:
equivalent sampling (Fluke 123)
.....
20 ns to 500 ns/div
equivalent sampling (Fluke 124)
.....
10 ns to 500 ns/div
real time sampling
...................................
1
µ
s to 5 s/div
Single (real time)
.....................................
1
µ
s to 5 s/div
Roll
(real time)
.........................................
1s to 60 s/div
Sampling Rate (for both channels simultaneously)
Equivalent sampling (repetitive signals)
........................
up to 1.25 GS/s
Real time sampling:
1
µ
s to 5 ms/div
..........................................
25 MS/s
10 ms to 60 s/div
..........................................
5 MS/s
Time Base Accuracy
Equivalent sampling
..................
±
(0.4% +0.04 time/div)
Real time sampling
....................
±
(0.1% +0.04 time/div)
Glitch Detection
..................
40 ns @
20 ns to 5 ms/div
200 ns @ 10 ms to 60 s/div
Glitch detection is always active.
Horizontal Move
...........................................
10 divisions
Trigger point can be positioned anywhere across the
screen.
Trigger
Screen Update
..............................
Free Run, On Trigger
Source
..............................................................
A, B, EXT
EXTernal via optically isolated trigger probe ITP120
(optional accessory)
Sensitivity A and B (Fluke 123)
@ DC to 5 MHz
.............................
0.5 divisions or 5 mV
@ 25 MHz
..................................................
1.5 divisions
@ 40 MHz
.....................................................
4 divisions
Sensitivity A and B (Fluke 124)
@ DC to 5 MHz
.............................
0.5 divisions or 5 mV
@ 40 MHz
..................................................
1.5 divisions
@ 60 MHz
.....................................................
4 divisions
Slope
...................................................
Positive, Negative
Video on A
..........................
interlaced video signals only
Modes
................................................
Lines, Line Select
Standards
..........................
PAL, NTSC, PAL+, SECAM
Polarity
..............................................
Positive, Negative
Sensitivity
.........................................
0.6 divisions sync.