Fluke 124 FE 123 & 124 Users Manual - Page 75

Input A, Continuity CONT, Diode, Capacitance CAP

Page 75 highlights

Input A Ohm (Ω) Ranges ....... 500Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ Accuracy 0.6% +5 counts) Full Scale Reading: 500Ω to 5 M 5000 counts 30 M 3000 counts Measurement Current 0.5 mA to 50 nA decreases with increasing ranges Open Circuit Voltage 4V Continuity (CONT) Beep 30Ω ±5Ω) in 50Ω range Measurement Current 0.5 mA Detection of shorts of 1 ms Diode Measurement Voltage: @0.5 mA 2.8V @open circuit 4V Accuracy 2% +5 counts) Measurement Current 0.5 mA Polarity on input A, - on COM Capacitance (CAP) Ranges 50 nF, 500 nF, 5 µF, 50 µF, 500 µF Accuracy 2% +10 counts) Full Scale Reading 5000 counts 4 Specifications Dual Input Meter Measurement Current 5 µA to 0.5 mA increases with increasing ranges Dual slope integrating measurement with parasitic serial and parallel resistance cancellation. 67

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86

Specifications
Dual Input Meter
4
67
Input A
Ohm (
)
Ranges
.......
500
, 5 k
, 50 k
, 500 k
, 5 M
, 30 M
Accuracy:
.........................................
±
(0.6% +5 counts)
Full Scale Reading:
500
to 5 M
.........................................
5000 counts
30 M
.....................................................
3000 counts
Measurement Current
.........................
0.5 mA to 50 nA
decreases with increasing ranges
Open Circuit Voltage
...............................................
<4V
Continuity (CONT)
Beep
.....................................
<(30
±
5
) in 50
range
Measurement Current
........................................
0.5 mA
Detection of shorts of
.........................................
1 ms
Diode
Measurement Voltage:
@0.5 mA
...........................................................
>2.8V
@open circuit
.......................................................
<4V
Accuracy
.............................................
±
(2% +5 counts)
Measurement Current
........................................
0.5 mA
Polarity
.....................................
+ on input A, - on COM
Capacitance (CAP)
Ranges
.................
50 nF, 500 nF, 5
µ
F, 50
µ
F,
500
µ
F
Accuracy
...........................................
±
(2% +10 counts)
Full Scale Reading
....................................
5000 counts
Measurement Current
...............................
5
µ
A to 0.5 mA
increases with increasing ranges
Dual slope integrating measurement with parasitic