Fluke 124 FE 123 & 124 Users Manual - Page 78

Environmental, Safety, Max. Input Voltage Input A and B

Page 78 highlights

Fluke 123/124 Users Manual Environmental Environmental MIL-PRF-28800F, Class 2 Temperature Operating 0 to 50 °C (32 to 122 °F) Storage 20 to 60 °C (-4 to 140 °F) Humidity Operating: @0 to 10 °C (32 to 50 °F noncondensing @10 to 30 °C (50 to 86 °F 95% @30 to 40 °C (86 to 104 °F 75% @40 to 50 °C (104 to 122 °F 45% Storage: @-20 to 60 °C (-4 to 140 °F noncondensing Altitude Operating 4.5 km (15 000 feet) Max. Input and Floating Voltage 600 Vrms upto 2 km, linearly derating to 400 Vrms @ 4.5 km Storage 12 km (40 000 feet) Vibration (sinusoidal MIL28800F, class 2, 3.8.4.2, 4.5.5.3.1: Max. 3g Shock MIL28800F, class 2, 3.8.5.1, 4.5.5.4.1: Max. 30g 70 Electromagnetic Compatibility (EMC) Emission EN 50081-1 (1992): EN55022 and EN60555-2 Immunity EN 50082-2 (1992): IEC1000-4-2, -3, -4, -5 (See also Tables 1 to 3) Enclosure Protection IP51, ref: IEC529 Safety Designed for measurements on 600 Vrms Category III Installations, Pollution Degree 2, per: • ANSI/ISA S82.01-1994 • EN61010-1 (1993) (IEC1010-1) • CAN/CSA-C22.2 No.1010.1-92 (including approval) • UL3111-1 (including approval) Max. Input Voltage Input A and B Direct on input or with leads 600 Vrms for derating, see Figure 4-1. With Banana-to BNC Adapter BB120 300 Vrms for derating, see Figure 4-1. Max. Floating Voltage from any terminal to ground 600 Vrms up to 400 Hz

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86

Fluke 123/124
Users Manual
70
Environmental
Environmental
.......................
MIL-PRF-28800F, Class 2
Temperature
Operating
................................
0 to 50
°
C (32 to 122
°
F)
Storage
.................................
-20 to 60
°
C (-4 to 140
°
F)
Humidity
Operating:
@0 to 10
°
C
(32 to 50
°
F)
.................
noncondensing
@10 to 30
°
C
(50 to 86
°
F)
................................
95%
@30 to 40
°
C (86 to 104
°
F)
................................
75%
@40 to 50
°
C (104 to 122
°
F)
..............................
45%
Storage:
@-20 to 60
°
C (-4 to 140
°
F)
..............
noncondensing
Altitude
Operating
......................................
4.5 km (15 000 feet)
Max. Input and Floating Voltage 600 Vrms upto 2 km,
linearly derating to 400 Vrms @ 4.5 km
Storage
...........................................
12 km (40 000 feet)
Vibration (sinusoidal)
....................................................
...........
MIL28800F, class 2, 3.8.4.2, 4.5.5.3.1: Max. 3g
Shock
...............................................................................
.........
MIL28800F, class 2, 3.8.5.1, 4.5.5.4.1: Max. 30g
Electromagnetic Compatibility (EMC)
Emission
.........................................
EN 50081-1 (1992):
EN55022 and EN60555-2
Immunity
.........................................
EN 50082-2 (1992):
IEC1000-4-2, -3, -4, -5
(See also Tables 1 to 3)
Enclosure Protection
..........................
IP51, ref: IEC529
Safety
Designed for measurements on
600 Vrms Category III
Installations, Pollution Degree 2, per:
ANSI/ISA S82.01-1994
EN61010-1 (1993) (IEC1010-1)
CAN/CSA-C22.2 No.1010.1-92 (including approval)
UL3111-1 (including approval)
Max. Input Voltage Input A and B
Direct on input or with leads
..........................
600 Vrms
for derating, see Figure 4-1.
With Banana-to BNC Adapter BB120
.............
300 Vrms
for derating, see Figure 4-1.
Max. Floating Voltage
from any terminal to ground
...........................
600 Vrms
up to 400 Hz