Fluke 124 FE 123 & 124 Users Manual - Page 74
Duty Cycle DUTY, Pulse Width PULSE, Amperes AMP, Temperature TEMP, Decibel dB, Crest Factor CREST,
View all Fluke 124 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 74 highlights
Fluke 123/124 Users Manual Accuracy: @1 Hz to 1 MHz 0.5% +2 counts) @1 to 10 MHz 1.0% +2 counts) @10 to 50 MHz (Fluke 123 2.5% +2 counts) @10 to 70 MHz (Fluke 124 2.5% +2 counts) (50 MHz in Autorange) Full Scale Reading 10 000 counts Duty Cycle (DUTY) Range 2% to 98% Frequency Range in Continuous Autoset 15 Hz (1 Hz) to 30 MHz Accuracy (Logic or Pulse waveforms): @1Hz to 1 MHz 0.5% +2 counts) @1 MHz to 10 MHz 1.0% +2 counts) Pulse Width (PULSE) Frequency Range in Continuous Autoset 15 Hz (1 Hz) to 30 MHz Accuracy (Logic or Pulse waveforms): @1 Hz to 1 MHz 0.5% +2 counts) @1 MHz to 10 MHz 1.0% +2 counts) Full Scale Reading 1000 counts Amperes (AMP with optional current probe Ranges......... same as VDC, VAC, VAC+DC, or PEAK Scale Factor ... 1 mV/A, 10 mV/A, 100 mV/A, and 1 V/A 66 Accuracy ......... same as VDC, VAC, VAC+DC, or PEAK (add current probe uncertainty) Temperature (TEMP) ... with optional temperature probe Range 200 °C/div (200 °F/div) Scale Factor 1 mV/°C and 1 mV/°F Accuracy........... as VDC (add temp. probe uncertainty) Decibel (dB) 0 dBV 1V 0 dBm (600Ω /50 1 mW referenced to 600Ω or 50Ω dB on VDC, VAC, or VAC+DC Full Scale Reading 1000 counts Crest Factor (CREST) Range 1 to 10 Accuracy 5% +1 count) Full Scale Reading 100 counts Phase Modes A to B, B to A Range 0 to 359 degrees Accuracy up to 1 MHz 2 degrees Accuracy 1 MHz to 5 MHz 5 degrees Resolution 1 degree