Invacare M91R Owners Manual - Page 108

Service Indicator Light

Page 108 highlights

9 TROUBLESHOOTING Service Indicator Light Diagnostics  For this procedure, refer to FIGURE 1. Service Indicator Light FIGURE 1 Service Indicator Light Diagnostics NUMBER OF FLASHES 1 2 3 4 5 6 ERROR CODE DESCRIPTION User Fault Battery Fault Left Motor Fault Right Motor Fault Left Park Brake Fault Right Park Brake Fault POSSIBLE SOLUTION Release joystick to neutral and try again. Charge the batteries. Refer to Charging Batteries on page 57. Check that battery cables are connected properly. If necessary, replace batteries. Refer to Replacing the Batteries and/or Battery Cables on page 90. Contact Invacare/Dealer for service. Contact Invacare/Dealer for service. Ensure brake lever is in the drive position before turning on the wheelchair. Ensure motor cable is plugged into the controller. Contact Invacare/Dealer for service. Ensure brake lever is in the drive position before turning on the wheelchair. Ensure motor cable is plugged into the controller. Contact Invacare/Dealer for service. Invacare® Pronto® M91™ 108 Part No 1143153

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116

9
TROUBLESHOOTING
Invacare® Pronto® M91™
108
Part No 1143153
Service Indicator Light Diagnostics
FIGURE 1
Service Indicator Light Diagnostics
For this procedure, refer to FIGURE 1.
NUMBER OF
FLASHES
ERROR CODE
DESCRIPTION
POSSIBLE SOLUTION
1
User Fault
Release joystick to neutral and try again.
2
Battery Fault
Charge the batteries. Refer to Charging Batteries
on page 57.
Check that battery cables are connected properly. If necessary, replace batteries. Refer to
Replacing the Batteries and/or Battery Cables
on page 90.
3
Left Motor Fault
Contact Invacare/Dealer for service.
4
Right Motor Fault
Contact Invacare/Dealer for service.
5
Left Park Brake Fault
Ensure brake lever is in the drive position before turning on the wheelchair. Ensure motor cable is
plugged into the controller.
Contact Invacare/Dealer for service.
6
Right Park Brake Fault
Ensure brake lever is in the drive position before turning on the wheelchair. Ensure motor cable is
plugged into the controller.
Contact Invacare/Dealer for service.
Service Indicator Light