Canon DP-V1710 User Manual - Page 29

Place the universal measuring probe pointing at the center of the video display.

Page 29 highlights

Operating the Video Display 2 Select [User 1] to [User 7] and then open the [Calibration] menu. 1 [Adjustment] [Picture Mode] [User 1] to [User 7] 2 [Adjustment] [Detail Settings] [Calibration] • Set each target value. 3 Press the jog dial and select [Start]. 29 • Please follow the information indicated on the screen. 4 Initialize the sensor. • Set the mode dial of the Universal Measuring Probe to [0-CAL]. • Press the jog dial of the video display, select [OK], and execute initialization. 5 Place the universal measuring probe pointing at the center of the video display. • Set the mode dial of the universal measuring probe to [MEAS] and place the probe as shown below according to the displayed content. Press the jog dial of the video display, select [OK], and execute calibration. Main Unit Konica Minolta Universal Measuring Probe CA-310 support: CA-PU32, CA-PU35 CA-210 support: CA-PU12, CA-PU15 Konica Minolta Display Color Analyzer CA-310, CA-210 6 Finish calibration. • When the message [Calibration is completed]. is displayed, press the jog dial and select [OK]. • If the message [Calibration error.] is displayed. Calibration has been terminated due to an error. The main unit returns to the state before calibration. ( 99) • To cancel calibration Press the jog dial during calibration and select [Cancel]. The main unit returns to the state before calibration. Note • Perform matrix calibration of the display color analyzer prior to calibration. If calibration is performed without performing matrix calibration, an error may occur. Refer to the CA-310 and CA-210 instruction manual for the detail operation. • Warming-up is necessary to stabilize the brightness of the video display. Wait at least 10 minutes after turning on the power before calibration. • Perform calibration in a dark room so that no external light enters the sensor. If external light enters the sensor, low brightness characteristics cannot be calibrated correctly. • Due to the characteristic of LCD panel and individual difference of CA-310 and CA-210, the calibration results may differ.

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113

Operating the Video Display
29
2 Select [User 1] to [User 7] and then open the [Calibration] menu.
1
[Adjustment]
[Picture Mode]
[User 1] to [User 7]
2
[Adjustment]
[Detail Settings]
[Calibration]
Set each target value.
3 Press the jog dial and select [Start].
Please follow the information indicated on the screen.
4 Initialize the sensor.
Set the mode dial of the Universal Measuring Probe to [0-CAL].
Press the jog dial of the video display, select [OK], and execute initialization.
5 Place the universal measuring probe pointing at the center of the video display.
Set the mode dial of the universal measuring probe to [MEAS] and place the probe as shown below according to
the displayed content. Press the jog dial of the video display, select [OK], and execute calibration.
Konica Minolta Universal Measuring Probe
CA-310 support: CA-PU32, CA-PU35
CA-210 support: CA-PU12, CA-PU15
Konica Minolta
Display Color Analyzer CA-310, CA-210
Main Unit
6 Finish calibration.
When the message [Calibration is completed]. is displayed, press the jog dial and select [OK].
If the message [Calibration error.] is displayed.
Calibration has been terminated due to an error. The main unit returns to the state before calibration. (
99)
To cancel calibration
Press the jog dial during calibration and select [Cancel]. The main unit returns to the state before calibration.
Note
Perform matrix calibration of the display color analyzer prior to calibration. If calibration is performed without performing matrix
calibration, an error may occur. Refer to the CA-310 and CA-210 instruction manual for the detail operation.
Warming-up is necessary to stabilize the brightness of the video display. Wait at least 10 minutes after turning on the power before
calibration.
Perform calibration in a dark room so that no external light enters the sensor. If external light enters the sensor, low brightness
characteristics cannot be calibrated correctly.
Due to the characteristic of LCD panel and individual difference of CA-310 and CA-210, the calibration results may differ.