#SSSW 001 00000000 F-14-2 Do not change service data identified as "not used"; they are set as initial settings. 14.3.2 SSSW-SW01: 14.3.2.1 List of Functions Bit Function 0 service error code 1 not used 2 not used 3 not used 4 not used 5 not used 6 not used 7 not used T-14-1 1 output - 14.3.2.2 Detailed Discussions of Bit 0 Selects whether or not service error codes are output. When output is selected, service error codes is report. 14.3.3 SSSW-SW03 14.3.3.1 List of Functions Bit Function T-14-2 0 not used 1 not used 2 not used 3 not used 4 not used 5 not used 6 not used 7 tonal signal before CED signal transmission 1 transmit 14.3.3.2 Detailed Discussions of Bit 7 Use it to enable/disable transmission of a 1080-Hz tonal signal before transmission of the CED signal. Select 'transmit' if errors occur frequently because of an echo when reception is from overseas. Memo: Any of the following error code may be indicated because of an echo at time of reception ##0005, ##0101, ##0106, ##0107, ##0114, ##0200, ##0201, ##0790 0 not output - 0 do not transmit Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0 Chapter 14 0016-2002 0016-2003 0016-2004 0016-2005 14-7