Invacare TDXSP Owners Manual 3 - Page 19

Repair or Service Information - tdx sp power wheelchair

Page 19 highlights

3 SAFETY Repair or Service Information ƽ WARNING Except for programming, DO NOT service or adjust the wheelchair while occupied, unless otherwise noted. Before adjusting, repairing or servicing the wheelchair, ALWAYS turn the wheelchair power Off, otherwise, injury or damage may occur. Wheelchairs should be examined during maintenance for signs of corrosion (water exposure, incontinence, etc.). Electrical components damaged by corrosion should be replaced immediately. Wheelchairs that are used by incontinent users and/or are frequently exposed to water may require replacement of electrical components more frequently. DO NOT overtighten hardware attaching to the frame. This could cause damage to the frame tubing. Transport ready packages are not retrofittable to existing models and are not field serviceable. TDX® OR TDX SP, SR, SI, SC and SPREE WHEELCHAIR BASES ONLY A pinch point exists between walking beam/head tube cap and telescoping tube when at the lowest seat to floor height. POWERED SEATING SYSTEMS ONLY Pinch point may occur when returning the tilted seat to the full upright position. Make sure the hands and body of the occupant, attendants and bystanders are clear of all pinch points before returning the tilted seat to the full upright position. PTO PLUS ONLY If tilt function is not operating and is in the down (seated) position, PRIOR to servicing the device, refer to the seating system service manual, part number 1125031, or contact Invacare Technical Services for instructions on how to safely service the tilt system, actuator or spring assist module. Call 800-832-4707. ALWAYS reinstall the spring safety bracket AND tether when servicing the PTO Plus seating system is complete. These safety devices MUST be in place at all times, otherwise, injury or damage may occur. Part No 1143192 19 Invacare® Adjustable ASBA Seats

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80

3
SAFETY
Part No 1143192
19
Invacare® Adjustable ASBA Seats
Repair or Service Information
±
WARNING
Except for programming, DO NOT service or adjust the wheelchair while occupied, unless otherwise noted.
Before adjusting, repairing or servicing the wheelchair, ALWAYS turn the wheelchair power Off, otherwise, injury or
damage may occur.
Wheelchairs should be examined during maintenance for signs of corrosion (water exposure, incontinence, etc.). Electrical
components damaged by corrosion should be replaced immediately.
Wheelchairs that are used by incontinent users and/or are frequently exposed to water may require replacement of
electrical components more frequently.
DO NOT overtighten hardware attaching to the frame. This could cause damage to the frame tubing.
Transport ready packages are not retrofittable to existing models and are not field serviceable.
TDX
®
OR TDX SP, SR, SI, SC and SPREE WHEELCHAIR BASES ONLY
A pinch point exists between walking beam/head tube cap and telescoping tube when at the lowest seat to floor height.
POWERED SEATING SYSTEMS ONLY
Pinch point may occur when returning the tilted seat to the full upright position. Make sure the hands and body of the
occupant, attendants and bystanders are clear of all pinch points before returning the tilted seat to the full upright position.
PTO PLUS ONLY
If tilt function is not operating and is in the down (seated) position, PRIOR to servicing the device, refer to the seating
system service manual, part number 1125031, or contact Invacare Technical Services for instructions on how to safely
service the tilt system, actuator or spring assist module. Call 800-832-4707.
ALWAYS reinstall the spring safety bracket AND tether when servicing the PTO Plus seating system is complete. These
safety devices MUST be in place at all times, otherwise, injury or damage may occur.