Fluke 724 Fluke 724 Users Manual - Page 42

Calibrating a Transmitter, keys. Enter these parameters by pressing

Page 42 highlights

724 Users Manual Calibrating a Transmitter Use the measurement (upper display) and source (lower display) modes to calibrate a transmitter. The following example shows how to calibrate a temperature transmitter. Connect the calibrator to the instrument under test as shown in Figure 15. Proceed as follows to calibrate a transmitter: 1. Press K for current measurement with loop power. 2. Press T (lower display). If desired, continue pressing this key to select the desired thermocouple type. M 3. If necessary, press for SOURCE mode. 4. Set your zero and span parameters by pressing X and W keys. Enter these parameters by pressing and holding J and G. For more information on setting parameters, see "Setting 0 % and 100 %" earlier in this manual. 5. Perform test checks at 0-25-50-75-100 % points by pressing H or I. Adjust the transmitter as necessary. 32

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56

724
Users Manual
32
Calibrating a Transmitter
Use the measurement (upper display) and source (lower
display) modes to calibrate a transmitter. The following
example shows how to calibrate a temperature
transmitter.
Connect the calibrator to the instrument under test as
shown in Figure 15. Proceed as follows to calibrate a
transmitter:
1.
Press
K
for current measurement with loop power.
2.
Press
T
(lower display). If desired, continue
pressing this key to select the desired thermocouple
type.
3.
If necessary, press
M
for SOURCE mode.
4.
Set your zero and span parameters by pressing
X
and
W
keys. Enter these parameters by pressing and
holding
J
and
G
. For more information on
setting parameters, see
Setting 0 % and 100 %
earlier in this manual.
5.
Perform test checks at 0-25-50-75-100 % points by
pressing
H
or
I
. Adjust the transmitter as
necessary.