724 Users Manual Ohms Measurement Accuracy ± Ω Ohms Range 4-Wire 2- and 3-Wire* 0 to 400 Ω 0.1 0.15 400 to 1.5 kΩ 0.5 1.0 1.5 to 3.2 kΩ 1 1.5 Excitation Current: 0.2 mA Maximum input voltage: 30 V Temperature coefficient -10 °C to 18 °C, +28 °C to 55 °C: ± 0.005 % of range per °C * 2-wire: Does not include lead resistance. 3-wire: Assumes matched leads with a total resistance not exceeding 100 Ω. Ohms Source Excitation Current from Accuracy Ohms Range Measurement Device ± Ω 15 to 400 Ω 0.15 to 0.5 mA 0.15 15 to 400 Ω 0.5 to 2 mA 0.1 400 to 1.5 kΩ 0.05 to 0.8 mA 0.5 1.5 to 3.2 kΩ 0.05 to 0.4 mA 1 Resolution 15 to 400 Ω 400 to 3.2 kΩ 0.1 Ω 1 Ω Temperature coefficient -10 °C to 18 °C, +28 °C to 55 °C: ± 0.005 % of resistance range per °C 40

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56

724
Users Manual
40
Ohms Measurement
Accuracy
±
Ohms Range
4-Wire
2- and 3-Wire*
0 to 400
0.1
0.15
400 to 1.5 k
0.5
1.0
1.5 to 3.2 k
1
1.5
Excitation Current:
0.2 mA
Maximum input voltage:
30 V
Temperature coefficient -10
°
C to 18
°
C, +28
°
C to
55
°
C:
±
0.005 % of range per
°
C
*
2-wire: Does not include lead resistance.
3-wire: Assumes matched leads with a total resistance
not exceeding 100
.
Ohms Source
Ohms Range
Excitation Current from
Measurement Device
Accuracy
±
15
to 400
0.15 to 0.5 mA
0.15
15 to 400
0.5 to 2 mA
0.1
400 to 1.5 k
0.05 to 0.8 mA
0.5
1.5 to 3.2 k
0.05 to 0.4 mA
1
Resolution
15 to 400
0.1
400 to 3.2 k
1
Temperature coefficient -10
°
C to 18
°
C, +28
°
C to
55
°
C:
±
0.005 % of resistance range per
°
C