Fluke 724 Fluke 724 Users Manual - Page 53

Loop Power Supply, General Specifications - meter

Page 53 highlights

Temperature Calibrator Specifications Loop Power Supply Voltage: 24 V Maximum current: 22 mA Short circuit protected General Specifications Operating temperature Storage temperature Operating altitude Relative Humidity (% RH operating without condensation) Vibration Safety Power requirements Size Weight -10 °C to 55 °C - 20 °C to 71 °C 3000 meters above mean sea level 90 % (10 to 30 °C) 75 % (30 to 40 °C) 45 % (40 to 50 °C) 35 % (50 to 55 °C) uncontrolled < 10 °C Random, 2 g, 5 to 500 Hz EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 4 AA alkaline batteries 96 x 200 x 47 mm. (3.75 x 7.9 x 1.86 in) 650 gm (1 lb, 7 oz) 43

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56

Temperature Calibrator
Specifications
43
Loop Power Supply
Voltage: 24 V
Maximum current: 22 mA
Short circuit protected
General Specifications
Operating temperature
-10
°
C to 55
°
C
Storage temperature
- 20
°
C to 71
°
C
Operating altitude
3000 meters above mean sea level
Relative Humidity (% RH operating without condensation)
90 % (10 to 30
°
C)
75 % (30 to 40
°
C)
45 % (40 to 50
°
C)
35 % (50 to 55
°
C)
uncontrolled < 10
°
C
Vibration
Random, 2 g, 5 to 500 Hz
Safety
EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2
No 1010.1:1992
Power requirements
4 AA alkaline batteries
Size
96 x 200 x 47 mm. (3.75 x 7.9 x 1.86 in)
Weight
650 gm (1 lb, 7 oz)