Fluke 724 Fluke 724 Users Manual - Page 51
Millivolt Measurement and Source*, Temperature, Thermocouples, Range, Resolution, Accuracy - specifications
View all Fluke 724 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 51 highlights
Millivolt Measurement and Source* Range Resolution Accuracy -10 mV to 75 mV 0.01 mV ±(0.025 % + 1 count) Maximum input voltage: 30 V Temperature coefficient -10 °C to 18 °C, +28 °C to 55 °C: ±0.005 % of range per °C *Select this function by pressing T. The signal is available at the thermocouple miniplug connector. Temperature, Thermocouples Type Range Measure and Source Accuracies (ITS-90) J -200 to 0 °C 0 to 1200 °C K -200 to 0 °C 0 to 1370 °C T -200 to 0 °C 0 to 400 °C E -200 to 0 °C 0 to 950 °C 1.0 °C 0.7 °C 1.2 °C 0.8 °C 1.2 °C 0.8 °C 0.9 °C 0.7 °C Temperature Calibrator Specifications R -20 to 0 °C 0 to 500 °C 500 to 1750 °C S -20 to 0 °C 0 to 500 °C 500 to 1750 °C B 600 to 800 °C 800 to 1000 °C 1000 to 1800 °C L -200 to 0 °C 0 to 900 °C U -200 to 0 °C 0 to 400 °C N -200 to 0 °C 0 to 1300 °C 2.5 °C 1.8 °C 1.4 °C 2.5 °C 1.8 °C 1.5 °C 2.2 °C 1.8 °C 1.4 °C 0.85 °C 0.7 °C 1.1 °C 0.75 °C 1.5 °C 0.9 °C Resolution: J, K, T, E, L, N, U: B, R, S: 0.1 °C, 0.1 °F 1 °C, 1 °F 41