Fluke 345 FE 345 Users Manual - Page 18
Specifications, Electrical Data
![]() |
View all Fluke 345 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 18 highlights
345 Users Manual Specifications Electrical Data All accuracies specified at 23 °C ± 1 °C Temperature coefficient of current ≤ ±0.15 % of rdg per °C Temperature coefficient of voltage ≤ ±0.15 % of rdg per °C Current Measurement (DC, DC RMS, AC RMS) Measuring range 0 - 2000 A dc, 1400 ac rms Autorange facility 40 A / 400 A / 2000 A Resolution 10 mA in 40 A range 100 mA in 400 A range 1 A in 2000 A range Accuracy RMS and DC I > 10 A 1.5 % rdg ± 5 digits I < 10 A 0.2 A AVE I > 10 A 3 % rdg ± 5 digits I < 10 A 0.5 A Pk I > 10 A 5 % rdg ± 5 digits I < 10 A 0.5 A AHr I > 10 AHr 2 % rdg ± 5 digits I < 10 AHr 0.5 AHr CF (Crest Factor) 1.1 ≤ CF < 3 3 % rdg ± 5 digits 3 ≤ CF < 5 5 % rdg ± 5 digits Resolution 0.01 RPL (Ripple) 2 % ≤ RPL< 100 3 % rdg ± 5 digits 100 % ≤ RPL< 600 5 % rdg ± 5 digits Resolution 0.1 % IDC > 5 A, IAC > 2 A All measurements DC and 15 Hz to 1 kHz. Maximum overload 10,000 A or rms x frequency < 400,000. Amps rms is a true rms measurement (ac + dc) 4
![](/manual_guide/products/fluke-345-fe-345-users-manual-862f3a3/18.png)