Campbell Scientific CSAT3 CSAT3 3-D Sonic Anemometer - Page 19
CSAT3 Outputs
View all Campbell Scientific CSAT3 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 19 highlights
CSAT3 Three Dimensional Sonic Anemometer TABLE 5-8. FW05/FWC-L35 Fine Wire Thermocouple Description Signal High Signal Low Color Purple Red 6. CSAT3 Outputs The CSAT3 has three signal outputs: SDM, RS-232, and analog. Campbell Scientific recommends that the SDM output be used when ever possible, followed by RS-232, and finally analog. TABLE 6-1 summarizes the difference between the three output options. In general SDM communications has the following advantages: lowest CSAT3 current drain, best control and diagnostics, and highest data resolution. TABLE 6-1. CSAT3 Outputs Typical Current Drain @ 10 Hz External Trigger Source Output Dynamic Range Diagnostics Available SDM 51 mA @ 12 Vdc yes 19 bit yes RS-232 89 mA @ 12 Vdc yes 19 bit yes Analog 99 mA @ 12 Vdc no 12 bit no 6.1 SDM Output SDM is a Campbell Scientific communication protocol used between a Campbell Scientific datalogger and a smart sensor. TABLE 6-2 lists all the Campbell Scientific dataloggers that support SDM communications, along with the required datalogger OS/PROM and CSAT3 embedded code. 11