CSAT3 Three Dimensional Sonic Anemometer FIGURE 11-4. Recommended settings for the wind vertical axis FIGURE 11-5. Recommended settings for the wind vertical axis labels Collect and graph 1 minute of wind data from the CSAT3 while it is in the zero wind environment (FIGURE 11-6). The wind offset should be less than ± 4 cm s-1 (0.04 m s-1) for ux and uy, and less than ± 2 cm s-1 (0.02 m s-1) for uz. If the CSAT3 wind offset is greater than these specifications, contact Campbell Scientific, Inc. for an RMA number to recalibrate the CSAT3. 41