Fluke 225C User Manual - Page 130

Meter, DMM Measurements on Meter Inputs, Meter Input, Meter Functions, General, Continuity CONT

Page 130 highlights

Fluke 19xC-2x5C Users Manual Meter Meter Input Input Coupling DC Frequency Response DC to 10 kHz (-3 dB) Input Impedance 1 MΩ (±1 %)//10 pF (±1.5 pF) Max. Input Voltage 1000 V CAT II 600 V CAT III (For detailed specifications, see "Safety") Meter Functions Ranging Auto, Manual Modes Normal, Relative 120 DMM Measurements on Meter Inputs The accuracy of all measurements is within ± (% of reading + number of counts) from 18 °C to 28 °C. Add 0.1x (specific accuracy) for each °C below 18 °C or above 28 °C. General DC Common Mode Rejection (CMRR 100 dB AC Common Mode Rejection at 50, 60, or 400 Hz ......>60 dB Ohms (Ω) Ranges 500.0 Ω, 5.000 kΩ, 50.00 kΩ, 500.0 kΩ, 5.000 MΩ, 30.00 MΩ Full Scale Reading 500 Ω to 5 M 5000 counts 30 M 3000 counts Accuracy 0.6 % +5 counts) Measurement Current 0.5 mA to 50 nA, ±20 % decreases with increasing ranges Open Circuit Voltage 4 V Continuity (CONT) Beep 50 Ω (±30 Ω) Measurement Current 0.5 mA, ±20 % Detection of shorts of 1 ms

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138
  • 139
  • 140
  • 141
  • 142
  • 143
  • 144
  • 145
  • 146
  • 147
  • 148
  • 149
  • 150
  • 151
  • 152
  • 153
  • 154
  • 155
  • 156
  • 157
  • 158
  • 159

Fluke 19xC-2x5C
Users Manual
120
Meter
Meter Input
Input Coupling
..............................................................
DC
Frequency Response
.......................
DC to 10 kHz (-3 dB)
Input Impedance
................
1 M
Ω
(±1 %)//10 pF (±1.5 pF)
Max. Input Voltage
..............................
1000 V CAT II
600 V CAT III
(For detailed specifications, see “Safety”)
Meter Functions
Ranging
.......................................................
Auto, Manual
Modes
.....................................................
Normal, Relative
DMM Measurements on Meter Inputs
The accuracy of all measurements is within
±
(% of
reading + number of counts) from 18
°
C to 28
°
C.
Add 0.1x (specific accuracy) for each
°
C below 18
°
C or
above 28
°
C.
General
DC Common Mode Rejection (CMRR)
..................
>100 dB
AC Common Mode Rejection at 50, 60, or 400 Hz
......
>60 dB
Ohms (
Ω
)
Ranges
..................................
500.0
Ω
, 5.000 k
Ω
, 50.00 k
Ω
,
500.0 k
Ω
, 5.000 M
Ω
, 30.00 M
Ω
Full Scale Reading
500
Ω
to 5 M
Ω
...............................................
5000 counts
30 M
Ω
...........................................................
3000 counts
Accuracy
...............................................
±
(0.6 % +5 counts)
Measurement Current
...................
0.5 mA to 50 nA,
±
20 %
decreases with increasing ranges
Open Circuit Voltage
....................................................
<4 V
Continuity (CONT)
Beep
...........................................................
<50
Ω
(
±
30
Ω
)
Measurement Current
..................................
0.5 mA,
±
20 %
Detection of shorts of
................................................
1 ms