Fluke 225C User Manual - Page 87
Profibus DP/RS-485, Profibus DP Measurement Connections
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 87 highlights
Profibus DP/RS-485 Default probe setting 10:1. Use the Fluke 10:1 probes. 1 Connect the red probe to the test tool input A and the grey probe to input B. 2 Connect the ground lead of the input A probe to the DP bus RxD/TxD N (-) 3 Connect the ground lead of the input B probe to the DP bus cable shield. 4 Connect the probe tip of both probes to the DP bus RxD/TxD P (+) DATA + DATA DATA GND 6 Using The Bushealth Function Input Connections and Tested Signals Table 9. Profibus DP Tested Signal Properties Signal Description V-Offset V-Level Data Rise Fall Jitter Signal Dist. Overshoot V offset Peak-peak voltage Bit width Rise time as % of bit width Fall time as % of bit width Jitter distortion Signal distortion (Manchester decoding, default setting) Overshoot (NRZ decoding, can be selected via limit setup) Note The bus normally has continuous data traffic. Figure 43. Profibus DP Measurement Connections 77