Fluke 225C User Manual - Page 139
Electromagnetic Immunity, No visible disturbance
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 139 highlights
10 Specifications Electromagnetic Immunity Electromagnetic Immunity The Fluke 19xC-2x5C series, including standard accessories, conforms with the EEC directive 2004/108/EC for EMC immunity, as defined by EN-61326-1:2006, with the addition of the following tables. Scope Mode (10 ms/div): Trace disturbance with VPS210 voltage probe shorted Table 12 No visible disturbance Frequency range 10 kHz to 20 MHz Frequency range 20 MHz to 100 MHz Frequency range 100 MHz to 1 GHz Frequency range 1.4 GHz to 2.0 GHz E = 3V/m 2 mV/div to 100 V/div 200 mV/div to 100 V/div 500 mV/div to 100 V/div *) 2 mV/div to 100 V/div (*) With the 20 MHz Bandwidth Filter switched on: no visible disturbance. With the 20 MHz Bandwidth Filter switched off: disturbance is max 2 div. Table 13 No visible disturbance Frequency range 2.0 GHz to 2.7 GHz E = 1V/m 2 mV/div to 100 V/div Table 14 Disturbance less than 10% of full scale Frequency range 20 MHz to 100 MHz E = 3V/m 10 mV/div to 100 mV/div Test Tool ranges not specified in tables 12, 13 and 14 may have a disturbance of more than 10% of full scale. 129