Fluke 225C User Manual - Page 140
Meter Mode Vdc, Vac, Vac+dc, Ohm and Continuity: Reading disturbance with test leads shorted
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 140 highlights
Fluke 19xC-2x5C Users Manual Meter Mode (Vdc, Vac, Vac+dc, Ohm and Continuity): Reading disturbance with test leads shorted Table 15 Disturbance less than 1% of full scale Frequency range 10 kHz to 1 GHz E = 3V/m 500 mV to 1000 V , 500 Ohm to 30 MOhm ranges No visible disturbance Frequency range 1.4 GHz to 2 GHz Table 16 E = 3V/m 500 mV to 1000 V, 500 Ohm to 30 MOhm ranges No visible disturbance Frequency range 2 GHz to 2.7 GHz Table 17 E = 1V/m 500 mV to 1000 V , 500 Ohm to 30 MOhm ranges 130
Fluke 19xC-2x5C
Users Manual
130
Meter Mode (Vdc, Vac, Vac+dc, Ohm and Continuity): Reading disturbance with test leads shorted
Table 15
Disturbance less than 1% of full scale
E = 3V/m
Frequency range 10 kHz to 1 GHz
500 mV to 1000 V , 500 Ohm to 30 MOhm ranges
Table 16
No visible disturbance
E = 3V/m
Frequency range 1.4 GHz to 2 GHz
500 mV to 1000 V, 500 Ohm to 30 MOhm ranges
Table 17
No visible disturbance
E = 1V/m
Frequency range 2 GHz to 2.7 GHz
500 mV to 1000 V , 500 Ohm to 30 MOhm ranges