Industrial ScopeMeter
Specifications
7
with iFlex clamp
Ranges
............................................................
20 A/division
Maximum Current
...........................................
75 A @ 40 Hz to 300 Hz
Frequency derating: I * F <22 500 A*Hz @ 300 Hz to 3000 Hz
Accuracy
.........................................................
±(1.5 % + 10 counts) @ 40 Hz to 60 Hz
±(3 % + 15 counts) @ 60 Hz to 1000 Hz
±(6 % + 15 counts) @ 1000 Hz to 3000 Hz
Temperature (TEMP) with optional temperature probe
Range
.................................................................
200
°
C/div (200
°
F/div)
Scale Factor
........................................................
1 mV/
°
C and 1 mV/
°
F
Accuracy
.............................................................
as VDC (add temp. probe uncertainty)
Decibel (dB)
0 dBV
..................................................................
1 V
0 dBm (600
Ω
/50
Ω
)
...........................................
1 mW referenced to 600
Ω
or 50
Ω
dB on
...................................................................
VDC, VAC, or VAC+DC
Full Scale Reading
..............................................
1000 counts
Crest Factor (CREST)
Range
.................................................................
1 to 10
Accuracy
....................................................................
±
(5 % + 1 count)
Full Scale Reading
..............................................
90 counts
Phase
Modes
.................................................................
A to B, B to A
Range
.................................................................
0 degrees to 359 degrees
Accuracy
<1 MHz
...........................................................
2 degrees
1 MHz to 5 MHz
..............................................
5 degrees
Resolution
...........................................................
1 degree
Power (125B)
Configurations
.....................................................
1 phase / 3 phase 3 conductor balanced loads (3 phase: fundamental component only,
AUTOSET mode only)
Power Factor (PF)
...............................................
ratio between Watts and VA
Range
.............................................................
0.00 to 1.00
Watt
.....................................................................
RMS reading of multiplying corresponding samples of input A (volts) and input B (amperes)
Full Scale reading
...........................................
999 counts
VA
.......................................................................
Vrms x Arms
Full Scale Reading
..........................................
999 counts
VA Reactive (VAR)
.............................................
((VA)
2
-W
2
)
Full Scale Reading
..........................................
999 counts
Vpwm
Purpose
...............................................................
to measure on pulse width modulated signals, like motor drive inverter outputs
Principle
..............................................................
readings show the effective voltage based on the average value of samples over a whole
number of periods of the fundamental frequency
Accuracy
.............................................................
as Vrms for sinewave signals
Input A
Ohm (
Ω
)
Ranges
125B
................................................................
50
, 500
Ω
, 5 k
Ω
, 50 k
Ω
, 500 k
Ω
, 5 M
Ω
, 30 M
Ω
124B, 123B
.....................................................
500
Ω
, 5 k
Ω
, 50 k
Ω
, 500 k
Ω
, 5 M
Ω
, 30 M
Ω
Accuracy
....................................................................
±
(0.6 % + 5 counts)
50
ω
±
(2 % + 20 counts)
Full Scale Reading
50
Ω
to 5 M
Ω
..................................................
5000 counts
30 M
Ω
.............................................................
3000 counts
Measurement Current
.........................................
0.5 mA to 50 nA, decreases with increasing ranges
Open Circuit Voltage
...........................................
<4 V