Fluke 125B/S Calibration Guide - Page 32
Input A and Input B AC Voltage Accuracy Test
View all Fluke 125B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 32 highlights
123B/124B/125B Calibration Manual Input A and Input B AC Voltage Accuracy Test XW Warning To prevent possible electrical shock, fire, or personal injury, make sure that the calibrator is in standby mode before making any connection between the calibrator and the test tool. Dangerous voltages will be present on the calibration source and connecting cables during the following steps. To test the Input A and Input B ac voltage accuracy: 1. Connect the test tool to the Calibrator. See Figure 6. 2. Setup V ac on the Test Tool: a. Push to select auto ranging (AUTO shows at the top of the display). Note For this procedure, do not press . b. Push . c. Push to open the MEASURE A menu. d. Select V ac. e. Push . f. Push . g. Push to open the MEASURE B menu. h. Select V ac. i. Push . 3. If A or B is not at the center, push to move them to the center and push to toggle between A or B. 4. For each sensitivity value listed in Table 5: a. On the Test Tool, push or to set the Input A and Input B sensitivity. The range is listed in the second column. b. Set up the Calibrator: i. Supply the AC voltage (NORMAL output, WAVE sine) listed in Table 5. ii. Push . c. Compare the Test Tool reading to the limit in Table 5. Table 5. Volts AC Measurement Verification Points Sensitivity (Oscilloscope) 200 mV/div 2 V/div 20V/div Range[1] (Meter) 500 mV 5 V 50 V Calibrator Output V rms Calibrator Frequency 500 mV 500 mV 5 V 5 V 50 V 50 V 60 Hz 20 kHz 20 kHz 60 Hz 60 Hz 20 kHz Limit A-B 494.0 to 506.0 486.0 to 514.0 4.860 to 5.140 4.940 to 5.060 49.40 to 50.60 48.60 to 51.40 24