Fluke 125B/S Calibration Guide - Page 47
Diode Zero, Gain, Table 16. Ohm Gain Calibration Points
View all Fluke 125B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 47 highlights
Industrial ScopeMeter Calibration Adjustment 13. Set the Calibrator to 0 V (zero). Table 16. Ohm Gain Calibration Points Gain Ohm (CL 0507) Gain Ohm (CL 0508) Gain Ohm (CL 0509) Gain Ohm (CL 0510) Gain Ohm (CL 0511) Gain Ohm (CL 0512) Gain Ohm (CL 0513) Calibration Step 50 Ω 0.5 kΩ 5 kΩ 50 kΩ 500 kΩ 5 MΩ 30 MΩ Input Value Diode Zero, Gain To do the Capacitance Gain calibration: 1. Push to select calibration adjustment step Diode Zero (CL 0520):IDLE. 2. Connect the Test Tool to the Calibrator as shown in Figure 13. A B 5522A CALIBRATOR Figure 13. Capacitance Gain Calibration Input Connections 3. Set up the Calibrator: a. Supply 0 mV DC. b. Push . 4. On the Test Tool: a. Push to start the calibration. b. Wait for Diode Zero (CL 0520):READY to show on the display. c. Push to select calibration adjustment step Gain Diode (CL 0521):IDLE. 5. Set the Calibrator to supply 1.25 V DC. 6. Push to start the calibration. 7. Wait until Gain Diode (CL 0521):READY shows on the display. 39