Fluke 125B/S Calibration Guide - Page 47

Diode Zero, Gain, Table 16. Ohm Gain Calibration Points

Page 47 highlights

Industrial ScopeMeter Calibration Adjustment 13. Set the Calibrator to 0 V (zero). Table 16. Ohm Gain Calibration Points Gain Ohm (CL 0507) Gain Ohm (CL 0508) Gain Ohm (CL 0509) Gain Ohm (CL 0510) Gain Ohm (CL 0511) Gain Ohm (CL 0512) Gain Ohm (CL 0513) Calibration Step 50 Ω 0.5 kΩ 5 kΩ 50 kΩ 500 kΩ 5 MΩ 30 MΩ Input Value Diode Zero, Gain To do the Capacitance Gain calibration: 1. Push  to select calibration adjustment step Diode Zero (CL 0520):IDLE. 2. Connect the Test Tool to the Calibrator as shown in Figure 13. A B 5522A CALIBRATOR Figure 13. Capacitance Gain Calibration Input Connections 3. Set up the Calibrator: a. Supply 0 mV DC. b. Push . 4. On the Test Tool: a. Push  to start the calibration. b. Wait for Diode Zero (CL 0520):READY to show on the display. c. Push  to select calibration adjustment step Gain Diode (CL 0521):IDLE. 5. Set the Calibrator to supply 1.25 V DC. 6. Push  to start the calibration. 7. Wait until Gain Diode (CL 0521):READY shows on the display. 39

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54

Industrial ScopeMeter
Calibration Adjustment
39
13.
Set the Calibrator to 0 V (zero).
Diode Zero, Gain
To do the Capacitance Gain calibration:
1.
Push
to select calibration adjustment step
Diode Zero (CL 0520):IDLE
.
2.
Connect the Test Tool to the Calibrator as shown in Figure 13.
Figure 13. Capacitance Gain Calibration Input Connections
3.
Set up the Calibrator:
a.
Supply 0 mV DC.
b.
Push
.
4.
On the Test Tool:
a.
Push
to start the calibration.
b.
Wait for
Diode Zero (CL 0520):READY
to show on the display.
c.
Push
to select calibration adjustment step
Gain Diode (CL 0521):IDLE
.
5.
Set the Calibrator to supply 1.25 V DC.
6.
Push
to start the calibration.
7.
Wait until
Gain Diode (CL 0521):READY
shows on the display.
Table 16. Ohm Gain Calibration Points
Calibration Step
Input Value
Gain Ohm (CL 0507)
50
Ω
Gain Ohm (CL 0508)
0.5 k
Ω
Gain Ohm (CL 0509)
5 k
Ω
Gain Ohm (CL 0510)
50 k
Ω
Gain Ohm (CL 0511)
500 k
Ω
Gain Ohm (CL 0512)
5 M
Ω
Gain Ohm (CL 0513)
30 M
Ω
5522A
CALIBRATOR
A
B