Fluke 125B/S Calibration Guide - Page 37
Table 9. V DC and V AC High Voltage Verification Points, Sensitivity, Oscilloscope, Range, Meter
View all Fluke 125B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 37 highlights
Industrial ScopeMeter Performance Verification 2. Setup V ac and V dc for Input A on the Test Tool: a. Push to select auto ranging (AUTO shows at the top of the display). b. Push . Note For this procedure, do not push . c. Push to open the INPUT A MEASUREMENTS menu. d. Select V ac. e. Push . f. Push to open the INPUT A MEASUREMENTS menu. g. Select V dc. V dc becomes the main reading, V ac is the secondary reading. 3. Setup V ac and V dc for Input B on the Test Tool: h. Push . i. Push to open the INPUT B MEASUREMENTS menu. j. Select V ac. k. Push . l. Push to open the INPUT B MEASUREMENTS menu. m. Select V dc. 4. Move the Input A and Input B ground level to the center grid line. 5. Push to switch between moving A or B. 6. If A or B is not at the center, push to move them to the center. 7. For each sensitivity value listed in Table 9: a. On the Test Tool, push or to set the Input A and Input B sensitivity. The range is listed in the second column of Table 9. b. Set up the Calibrator: i. Supply the AC voltage (NORMAL output, WAVE sine) listed in Table 9. ii. Push . c. Compare the main reading (V-dc) and secondary reading (V-ac) to the readings in Table 9. Table 9. V DC and V AC High Voltage Verification Points Sensitivity (Oscilloscope) Range (Meter) 200 V/div 500 V Calibrator Output V rms Calibrator Frequency 0 V +500 V -500 V 500 V 500 V DC DC DC 60 Hz 10 kHz Main (DC) Reading A-B -000.5 to +000.5 +497.0 to +503.0 -497.0 to -503.0 Secondary (AC) Reading A-B 494.0 to 506.0 486.0 to 514.0 29