Fluke 125B/S Calibration Guide - Page 50
Bail Removal, Battery Replacement/Charge
View all Fluke 125B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 50 highlights
123B/124B/125B Calibration Manual To avoid loss of data, do one of the following before you remove the battery pack: • Store the data on a computer or a USB device. • Connect the power adapter. To replace the battery pack: 1. Turn off the Test Tool. 2. Remove all probes and test leads. 3. Unlock the battery cover. See Figure 14. 4. Lift the battery cover and remove it from the Test Tool. 5. Lift one side of the battery pack and remove it from the Test Tool. 6. Install a good battery pack. 7. Place the battery cover into position and lock. 1 2 3 Figure 14. Battery Replacement/Charge Bail Removal To remove the bail: 1. With the bail open, press sideways on the bail until one of the pivots pops out. See Figure 14. 2. Remove the bail. To reinsert: 1. With the bail in the open position, insert one pivot and press in and over on the bail. 2. Insert the second pivot into place. 42 CHARGE