Fluke 125B/S Calibration Guide - Page 29

Set up the Test Tool to test Input B, In the Scope Volt mode, supply 1.1 V DC.

Page 29 highlights

Industrial ScopeMeter Performance Verification 16. Verify that traces do not show on the display of the Test Tool and that Waiting shows on the status line at the top of the display. If traces show on the display and Hold shows on the status line at the top of the display, push  to reset the Test Tool. 17. Use the EDIT FIELD function on the Calibrator to slowly decrease the voltage in 0.1 V steps until the Test Tool triggers and the traces show on the display. 18. Verify that the voltage on the Calibrator is between +0.5 V and +1.5 V when the Test Tool triggers. 19. To repeat the test, start at step 15. 20. Set the Calibrator to Standby. 21. Push to clear the display. 22. Set up the Test Tool to test Input B: a. Push . b. Push  to open the SCOPE SETTINGS menu. c. Make these selections: • Trigger Input: B • Update: Single • Type: Normal • Waveform: Normal 23. Push  to setup the Test Tool to test the positive slope. Use  to set the trigger level to +2 divisions. For positive slope triggering, the trigger level is the top of the trigger icon. 24. Set up the Calibrator: a. In the Scope Volt mode, supply 1.1 V DC. b. Push . 25. Verify that traces do not show on the display of the Test Tool and that Waiting shows on the status line at the top of the display. If traces show on the display and Hold shows on the status line at the top of the display, push  to reset the Test Tool. 26. Use the EDIT FIELD function on the Calibrator to slowly increase the voltage in 0.1 V steps until the Test Tool triggers and the traces show on the display. 27. Verify that the voltage on the Calibrator is between +1.5 V and +2.5 V when the Test Tool triggers. 28. To repeat the test, start at step 24. 29. Set the Calibrator to Standby. 30. Push to clear the display. 31. Set up the Test Tool to test the negative slope. Use  to set the trigger level to +1 divisions. For negative slope triggering, the trigger level is the bottom of the trigger icon. 32. Set up the Calibrator: a. Supply +2 V DC. b. Push . 33. Verify traces do not show on the display of the Test Tool and that Waiting shows on the status line at the top of the display. If traces show on the display and Hold shows on the status line at the top of the display, push  to reset the Test Tool. 34. Use the EDIT FIELD function on the Calibrator to slowly decrease the voltage in 0.1 V steps until the Test Tool triggers and the traces show on the display. 35. Verify that the voltage on the Calibrator is between +0.5 V and +1.5 V when the Test Tool triggers. 36. To repeat the test, start at step 32. 21

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54

Industrial ScopeMeter
Performance Verification
21
16.
Verify that traces do not show on the display of the Test Tool and that Waiting shows on the status line at the top
of the display.
If traces show on the display and Hold shows on the status line at the top of the display, push
to reset the
Test Tool.
17.
Use the EDIT FIELD function on the Calibrator to slowly decrease the voltage in 0.1 V steps until the Test Tool
triggers and the traces show on the display.
18.
Verify that the voltage on the Calibrator is between +0.5 V and +1.5 V when the Test Tool triggers.
19.
To repeat the test, start at step 15.
20.
Set the Calibrator to Standby.
21. Push
to clear the display.
22.
Set up the Test Tool to test Input B:
a.
Push
.
b.
Push
to open the SCOPE SETTINGS menu.
c.
Make these selections:
Trigger Input:
B
Update:
Single
Type:
Normal
Waveform:
Normal
23. Push
to setup the Test Tool to test the positive slope. Use

to
set the trigger level to +2 divisions.
For
positive slope
triggering, the trigger level is the
top
of the trigger icon.
24.
Set up the Calibrator:
a.
In the Scope Volt mode, supply 1.1 V DC.
b.
Push
.
25.
Verify that traces do not show on the display of the Test Tool and that Waiting shows on the status line at the top
of the display.
If traces show on the display and Hold shows on the status line at the top of the display, push
to reset the
Test Tool.
26.
Use the EDIT FIELD function on the Calibrator to slowly increase the voltage in 0.1 V steps until the Test Tool
triggers and the traces show on the display.
27.
Verify that the voltage on the Calibrator is between +1.5 V and +2.5 V when the Test Tool triggers.
28.
To repeat the test, start at step 24.
29.
Set the Calibrator to Standby.
30. Push
to clear the display.
31.
Set up the Test Tool to test the negative slope. Use

to
set the trigger level to +1 divisions.
For
negative
slope
triggering, the trigger level is the
bottom
of the trigger icon.
32.
Set up the Calibrator:
a.
Supply +2 V DC.
b.
Push
.
33.
Verify traces do not show on the display of the Test Tool and that Waiting shows on the status line at the top of
the display.
If traces show on the display and Hold shows on the status line at the top of the display, push
to reset the
Test Tool.
34.
Use the EDIT FIELD function on the Calibrator to slowly decrease the voltage in 0.1 V steps until the Test Tool
triggers and the traces show on the display.
35.
Verify that the voltage on the Calibrator is between +0.5 V and +1.5 V when the Test Tool triggers.
36.
To repeat the test, start at step 32.