Fluke 125B/S Calibration Guide - Page 29
Set up the Test Tool to test Input B, In the Scope Volt mode, supply 1.1 V DC.
View all Fluke 125B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 29 highlights
Industrial ScopeMeter Performance Verification 16. Verify that traces do not show on the display of the Test Tool and that Waiting shows on the status line at the top of the display. If traces show on the display and Hold shows on the status line at the top of the display, push to reset the Test Tool. 17. Use the EDIT FIELD function on the Calibrator to slowly decrease the voltage in 0.1 V steps until the Test Tool triggers and the traces show on the display. 18. Verify that the voltage on the Calibrator is between +0.5 V and +1.5 V when the Test Tool triggers. 19. To repeat the test, start at step 15. 20. Set the Calibrator to Standby. 21. Push to clear the display. 22. Set up the Test Tool to test Input B: a. Push . b. Push to open the SCOPE SETTINGS menu. c. Make these selections: • Trigger Input: B • Update: Single • Type: Normal • Waveform: Normal 23. Push to setup the Test Tool to test the positive slope. Use to set the trigger level to +2 divisions. For positive slope triggering, the trigger level is the top of the trigger icon. 24. Set up the Calibrator: a. In the Scope Volt mode, supply 1.1 V DC. b. Push . 25. Verify that traces do not show on the display of the Test Tool and that Waiting shows on the status line at the top of the display. If traces show on the display and Hold shows on the status line at the top of the display, push to reset the Test Tool. 26. Use the EDIT FIELD function on the Calibrator to slowly increase the voltage in 0.1 V steps until the Test Tool triggers and the traces show on the display. 27. Verify that the voltage on the Calibrator is between +1.5 V and +2.5 V when the Test Tool triggers. 28. To repeat the test, start at step 24. 29. Set the Calibrator to Standby. 30. Push to clear the display. 31. Set up the Test Tool to test the negative slope. Use to set the trigger level to +1 divisions. For negative slope triggering, the trigger level is the bottom of the trigger icon. 32. Set up the Calibrator: a. Supply +2 V DC. b. Push . 33. Verify traces do not show on the display of the Test Tool and that Waiting shows on the status line at the top of the display. If traces show on the display and Hold shows on the status line at the top of the display, push to reset the Test Tool. 34. Use the EDIT FIELD function on the Calibrator to slowly decrease the voltage in 0.1 V steps until the Test Tool triggers and the traces show on the display. 35. Verify that the voltage on the Calibrator is between +0.5 V and +1.5 V when the Test Tool triggers. 36. To repeat the test, start at step 32. 21