Fluke 125B/S Calibration Guide - Page 30
Input A and Input B DC Voltage Accuracy Test
View all Fluke 125B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 30 highlights
123B/124B/125B Calibration Manual Input A and Input B DC Voltage Accuracy Test XW Warning To prevent possible electrical shock, fire, or personal injury, make sure that the calibrator is in standby mode before making any connection between the calibrator and the test tool. Dangerous voltages will be present on the calibration source and connecting cables during the following steps. To test the Input A and B voltage accuracy: 1. Connect the test tool to the Calibrator. See Figure 6. A B 5522A CALIBRATOR BB120 BNC PM9091/001 (1.5 m) PM9092/001 (0.5 m) PM9081 PM9093 Figure 6. Test Tool Input A-B to Calibrator Normal Output 2. Setup V dc on the Test Tool: a. Push . b. Push . c. Push to open the MEASURE A menu. d. Select V dc. e. Push . f. Push . g. Push to open the MEASURE B menu. h. Select V dc. i. Push . 3. Use to change the time base to select manual time base ranging and lock the time base on 10 ms/div. 4. Move the Input A and Input B ground level to the center grid line. 5. Push to toggle between Move A and Move B. If A or B is not at the center, use to move them to the center. 6. Set up the input settings on the Test Tool: a. Push . b. Push to open the INPUT SETTINGS A menu. c. Make these selections: • Select Coupling: DC • Readings: Smooth • Noise filter: ON. 22