Fluke 125B/S Calibration Guide - Page 25
Input B Frequency Measurement Accuracy Test, Connect the Test Tool to the Calibrator. See
View all Fluke 125B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 25 highlights
Input B Frequency Measurement Accuracy Test To test the Input B frequency measurement accuracy: 1. Connect the Test Tool to the Calibrator. See Figure 4. A B BNC Industrial ScopeMeter Performance Verification 5522A CALIBRATOR BB120 50Ω BNC Figure 4. Test Tool Input B to Calibrator Scope Output 50 Ω 2. Set up the Test Tool: a. Push to select auto ranging (AUTO shows at the top of the display). b. Push . c. Push to open the INPUT B MEASUREMENTS menu. d. Select Hz. e. Push twice to close the menu. f. Push . g. Push to open the SCOPE SETTINGS menu. h. Make these selections: • Trigger Input: B • Update: Free Run • Type: Normal • Waveform: Normal 3. Set up the Calibrator: a. Supply a leveled sine wave of 600 mV peak-to-peak (SCOPE output, MODE levsine). b. Push . 4. For each Output in Table 3: a. On the Calibrator, supply the frequency listed. Start with 1 MHz. b. Compare the Input B main reading on the Test Tool with the reading range listed in Table 3. 17