Fluke 125B/S Calibration Guide - Page 38
Resistance Measurements Test, Continuity Function Test
View all Fluke 125B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 38 highlights
123B/124B/125B Calibration Manual Resistance Measurements Test To test the resistance measurements: 1. Connect the Test Tool to the Calibrator. See Figure 9. A B 5522A CALIBRATOR Figure 9. Test Tool Input A to Calibrator Normal Output 4-Wire 2. Setup the Test Tool: a. Push to select auto ranging (AUTO shows at the top of the display). b. Push . c. Push to open the INPUT A MEASUREMENTS menu. d. Select OHM. 3. For each output listed in Table 10, set up the Calibrator: a. Supply the Output value. b. Push . c. Use the "COMP 2 wire" mode. d. Compare the main reading to the limit in Table 10. Table 10. Resistance Measurement Verification Points 0 Ω 400 Ω 4 kΩ 40 kΩ 400 kΩ 4 MΩ 30 MΩ Calibrator Output 000.0 to 000.5 397.1 to 402.9 3.971 to 4.029 39.71 to 40.29 397.1 to 402.9 3.971 to 4.029 29.77 to 30.23 Limit Continuity Function Test To test the continuity function: 1. Connect the Test Tool to the Calibrator. See Figure 9. 2. Set up the Test Tool: a. Push to select auto ranging (AUTO shows at the top of the display). b. Push . c. Push to open the INPUT A MEASUREMENTS menu. d. Select CONT. 30