Fluke 28IIEX Fluke 28IIEx User Manual - Page 17

Ex-Certification Data

Page 17 highlights

Ex-Certification Data • Ex-Type certificate no: • Ex-Designation: • Power Supply: • CE: CE0102 • Operating Temperature: -15 °C to 50 °C • Storage Temperature: -55 °C to +60 °C • Batteries: 3 AAA Alkaline batteries, NEDA 24A IEC LR03. Table 9 shows the approved batteries for this Product. For connections to intrinsically-safe circuits, observe these Product connections: Voltage - measurement input Ui = 65 V: U0 = 9.54 V C0 = 3.6 μF I0 = 3.7 mA L0 = 1000 mH P0 = 3.4 mW Ui = 65 V Ci = negligible Ii = negligible Li = negligible Current - measurement input Ii = 5 A: U0 = 0 V C0 = 1000 μF I0 = 9.7 μA L0 = 1000 mH P0 = 0 mWH Ui = 65 V Ci = negligible Ii = 5 A Li = negligible True-rms Digital Multimeter EX Safety Information mA/μA Jack U0 = 1.94 V Ui = 65 V C0 = 1000 μF Ci = negligible I0 = 9.7 μA Ii = Internally limited by a 440 mA fuse L0 = 1000 mH Li = negligible P0 = negligible For measurements on protected electric circuits: • Approved for Zones 2 and 1, device group II, explosion group IIC (explosive gases, vapors and mist), temperature class T4. • Approved for Zones 21 and 22, device group II, explosion group IIIC, conducting and non-conducting dust, fibers, and flyings. • Approved for use in mines. Device group I, explosion group I, methane, and coal dust. 7

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64

True-rms Digital Multimeter
EX Safety Information
7
Ex-Certification Data
Ex-Type certificate no:
Ex-Designation:
Power Supply:
CE: CE0102
Operating Temperature: -15
°
C to 50
°
C
Storage Temperature: -55
°
C to +60
°
C
Batteries: 3 AAA Alkaline batteries, NEDA 24A IEC
LR03. Table 9 shows the approved batteries for this
Product.
For connections to intrinsically-safe circuits, observe
these Product connections:
Voltage – measurement input U
i
= 65 V:
U
0
= 9.54 V
U
i
= 65 V
C
0
= 3.6
μ
F
C
i
= negligible
I
0
= 3.7 mA
I
i
= negligible
L
0
= 1000 mH L
i
= negligible
P
0
= 3.4 mW
Current – measurement input I
i
= 5 A:
U
0
= 0 V
U
i
= 65 V
C
0
= 1000
μ
F
C
i
= negligible
I
0
= 9.7
μ
A
I
i
= 5 A
L
0
= 1000 mH L
i
= negligible
P
0
= 0 mWH
mA/
μ
A Jack
U
0
= 1.94 V
U
i
= 65 V
C
0
= 1000
μ
F
C
i
= negligible
I
0
= 9.7
μ
A
I
i
= Internally limited by a 440 mA fuse
L
0
= 1000 mH L
i
= negligible
P
0
= negligible
For measurements on protected electric circuits:
Approved for Zones 2 and 1, device group II,
explosion group IIC (explosive gases, vapors and
mist), temperature class T4.
Approved for Zones 21 and 22, device group II,
explosion group IIIC, conducting and non-conducting
dust, fibers, and flyings.
Approved for use in mines. Device group I, explosion
group I, methane, and coal dust.