Add to My Manuals
Save this manual to your list of manuals |
Page 50 highlights
28 II Ex Users Manual Replace the batteries as follows, refer to Figure 11: 1. Turn the rotary switch to OFF and remove the test leads from the terminals. 2. Remove the six Torx-head screws from the case bottom and remove the battery door (). Note When you lift the battery door, make sure the rubber gasket stays attached to the battery compartment barrier. 3. Remove the three batteries and replace all three with AAA Alkaline batteries (). 4. Make sure the battery compartment gasket () is properly installed around the outside edge of the battery compartment barrier. 5. Align the battery compartment barrier with battery compartment while you replace the battery door. 6. Attach the door with the six Torx-head screws. Note It is recommended the batteries be removed from the Product for long periods of storage. Table 9. Approved Batteries Battery Description Manufacturer Duracell Procell MN2400 LR03 Duracell Plus MN2400 LR03 Duracell Max Tech No. 4703 Industrial Alcaline No. 4003 [1] Varta Eveready Energizer No. E92 Eveready Rayovac Alkaline AAA (U.S. Type) Rayovac Panasonic LR03XWA Panasonic [1] Minimum operating temperature is -10 °C. 40