Fluke 28IIEX Fluke 28IIEx User Manual - Page 59

DC Voltage, Conductance, and Resistance, Function, Range, Resolution, Accuracy, mV dc

Page 59 highlights

True-rms Digital Multimeter Detailed Specifications DC Voltage, Conductance, and Resistance Function Range Resolution Accuracy mV dc V dc 600.0 mV 6.000 V 60.00 V 600.0 V 1000 V 600.0 Ω 0.1 mV 0.001 V 0.01 V 0.1 V 1 V 0.1 Ω ±(0.1 % + 1) ±(0.05 % + 1) ±(0.2 % + 2) [2] 6.000 kΩ 60.00 kΩ Ω 600.0 kΩ 6.000 MΩ 50.00 MΩ 0.001 kΩ 0.01 kΩ 0.1 kΩ 0.001 MΩ 0.01 MΩ ±(0.2 % + 1) ±(0.6 % + 1) ±(1.0 % + 3) [1,3] nS 60.00 nS 0.01 nS ±(1.0 % + 10) [1,2,3] [1] Add 0.5 % of reading when measuring above 30 MΩ in the 50 MΩ range, and 20 counts below 33 nS in the 60 nS range. [2] When using the rel function to compensate for offsets. [3] >40 °C temperature coefficient is 0.1 x (specified accuracy)/°C. 49

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64

True-rms Digital Multimeter
Detailed Specifications
49
DC Voltage, Conductance, and Resistance
Function
Range
Resolution
Accuracy
mV dc
600.0 mV
0.1 mV
±
(0.1 % + 1)
V dc
6.000 V
0.001 V
±
(0.05 % + 1)
60.00 V
0.01 V
600.0 V
0.1 V
1000 V
1 V
Ω
600.0
Ω
0.1
Ω
±
(0.2 % + 2)
[2]
6.000 k
Ω
0.001 k
Ω
±
(0.2 % + 1)
60.00 k
Ω
0.01 k
Ω
600.0 k
Ω
0.1 k
Ω
±
(0.6 % + 1)
6.000 M
Ω
0.001 M
Ω
50.00 M
Ω
0.01 M
Ω
±
(1.0 % + 3)
[1,3]
nS
60.00 nS
0.01 nS
±
(1.0 % + 10)
[1,2,3]
[1]
Add 0.5 % of reading when measuring above 30 M
Ω
in the 50 M
Ω
range, and 20 counts below 33 nS in the 60 nS range.
[2]
When using the rel function to compensate for offsets.
[3]
>40
°
C temperature coefficient is 0.1 x (specified accuracy)/
°
C.