Fluke 8845A FE 8845A & 8846A Users Manual - Page 21
A Accuracy, Additional Errors, AC Voltage Specifications
View all Fluke 8845A manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 21 highlights
1 Introduction and Specifications Electrical Specifications 8846A Accuracy Accuracy is given as ± (% measurement + % of range) Range 100 mV 1 V 10 V 100 V 1000 V 24 Hour (23 ± 1 °C) 0.0025 + 0.003 0.0018 + 0.0006 0.0013 + 0.0004 0.0018 + 0.0006 0.0018 + 0.0006 90 Days (23 ± 5 °C) 0.0025 + 0.0035 0.0018 + 0.0007 0.0018 + 0.0005 0.0027 + 0.0006 0.0031 + 0.001 1 Year (23 ± 5 °C) 0.0037 + 0.0035 0.0025 + 0.0007 0.0024 + 0.0005 0.0038 + 0.0006 0.0041 + 0.001 Temperature Coefficient/ °C Outside 18 to 28 °C 0.0005 + 0.0005 0.0005 + 0.0001 0.0005 + 0.0001 0.0005 + 0.0001 0.0005 + 0.0001 8845A Accuracy Accuracy is given as ± (% measurement + % of range) Range 100 mV 1 V 10 V 100 V 1000 V 24 Hour (23 ± 1 °C) 0.003 + 0.003 0.002 + 0.0006 0.0015 + 0.0004 0.002 + 0.0006 0.002 + 0.0006 90 Days (23 ± 5 °C) 0.004 + 0.0035 0.003 + 0.0007 0.002 + 0.0005 0.0035 + 0.0006 0.0035 + 0.0010 1 Year (23 ± 5 °C) 0.005 + 0.0035 0.004 + 0.0007 0.0035 + 0.0005 0.0045 + 0.0006 0.0045 + 0.0010 Temperature Coefficient/ °C Outside 18 to 28 °C 0.0005 + 0.0005 0.0005 + 0.0001 0.0005 + 0.0001 0.0005 + 0.0001 0.0005 + 0.0001 Additional Errors Digits 6½ 6½ 5½ 5½ 4½ NPLC 100 10 1 .2 0.02 Additional Noise Error 0 % of range 0 % of range 0.001 % of range 0.001 % of range + 20 µV 0.01 % of range + 20 µV AC Voltage Specifications AC Voltage specifications are for ac sinewave signals >5 % of range. For inputs from 1 % to 5 % of range and